Smith B E, King P H, Schlain L
Department of Anesthesiology, Vanderbilt University Medical Center, Nashville, Tennessee.
Int J Clin Monit Comput. 1992;9(1):45-52. doi: 10.1007/BF01145901.
A clinical evaluation of the clinical utility, techniques of use, durability, accuracy, and potential complications of a newly available system for the continuous real-time, intra-arterial monitoring of arterial blood gas and acid base status (ABG) has been studied (Optex BioSentry System, Optex Biomedical, Incorporated, The Woodlands, Texas, U.S.A.). The system consists of three separate fiber optic channels with contained fluorescent and absorbant chemical analytes imbedded in a single probe and capable of insertion inside of a twenty gauge indwelling arterial catheter (The Optex Optode Sensor), along with an external monitor. The system was utilized during anesthesia and surgery in the care of five informed patients. Constant trend monitoring of all three variables was deemed satisfactory in four of the patients. The fifth sensor was damaged by a surgical assistant while in place and ceased to function. Comparison of Optode sensor readings with standard clinical laboratory ABG analysis was excellent in three uncomplicated patients (pH: bias -0.0183 pH units, precision: 0.0237 pH units; PCO2: bias 3.22 mmHg, precision 2.04 mmHg; PO2: bias -5.94 mmHg, precision 11.74 mmHg). Postoperative study suggested that discrepancies in a fourth patient may have been due to an incorrect 'offset' applied to the Optode sensor yielding a constant error.
对一种新上市的用于连续实时动脉内监测动脉血气和酸碱状态(ABG)的系统(美国德克萨斯州伍德兰市奥普泰克斯生物医学公司的奥普泰克斯生物哨兵系统)的临床实用性、使用技术、耐用性、准确性及潜在并发症进行了临床评估。该系统由三个独立的光纤通道组成,单个探头内包含荧光和吸收性化学分析物,能够插入20号留置动脉导管内(奥普泰克斯光极传感器),并配有一个外部监测器。在麻醉和手术过程中,对五名知情患者使用了该系统。四名患者对所有三个变量的持续趋势监测结果令人满意。第五个传感器在放置时被一名手术助手损坏,无法再使用。在三名未出现并发症的患者中,光极传感器读数与标准临床实验室ABG分析结果的对比情况良好(pH值:偏差 -0.0183 pH单位,精密度:0.0237 pH单位;PCO2:偏差3.22 mmHg,精密度2.04 mmHg;PO2:偏差 -5.94 mmHg,精密度11.74 mmHg)。术后研究表明,第四名患者出现差异可能是由于对光极传感器应用了错误的“偏移量”,从而产生了恒定误差。