Suppr超能文献

连续动脉内血气监测系统的多中心评估

Multisite evaluation of a continuous intraarterial blood gas monitoring system.

作者信息

Larson C P, Vender J, Seiver A

机构信息

Department of Anesthesia, Stanford University School of Medicine, California 94305.

出版信息

Anesthesiology. 1994 Sep;81(3):543-52. doi: 10.1097/00000542-199409000-00005.

Abstract

BACKGROUND

We compared the performance of a new, continuous intraarterial blood gas (CIABG) monitor with arterial values obtained periodically and analyzed by conventional equipment.

METHODS

A CIABG monitoring system consisting of a sterile, disposable, fiberoptic sensor and a microprocessor-controlled monitor with a self-contained calibration unit and detachable display panel was used. The sensor was inserted through a 20-G radial artery cannula. Light was transmitted from the monitor to the sensor tip where it reacted with fluorescent dyes sensitive to oxygen or hydrogen ions (analytes). The change in the intensity of the photoluminescent radiation caused by the analytes was measured every 20 s and derived blood gas values were displayed. Twenty-nine sensors were evaluated in 29 surgical or intensive care unit patients at one of three institutions (Stanford University Hospital, Evanston Memorial Hospital, and the Palo Alto Veterans Administration Hospital). The duration of study averaged 6 h (5-8 h) in the operating room, and 46 h (7-121 h) in the intensive care unit. A total of 552 values were compared with those obtained at regular intervals and analyzed in the hospital blood gas laboratory. Average bias (mean difference between lab value and CIABG), precision (SD of difference), and drift (change in the bias with time were determined.

RESULTS

At arterial oxygen tension (PO2) values of 32-528 mmHg, the average bias was -1% meaning that the average CIABG monitor values were 1% lower than those obtained by conventional equipment. The precision was 15%. At arterial PO2 values of 32-99 mmHg, average bias and precision were -0.3 +/- 8.9 mmHg. At arterial carbon dioxide tension (PCO2) values of 24-54 mmHg, average bias and precision were 1.3 +/- 3.3 mmHg, and at pHa values of 7.23-7.57, average bias and precision were 0.01 +/- 0.04. Observed drift per day was -1.2% for arterial PO2, 0.3 mmHg for arterial PCO2, and 0.01 for pH. Bias and precision for samples compared in two pairs of like-model in vitro blood gas analyzers were 0.4 +/- 4.6% for arterial PO2 over the full range, and 0.4 +/- 3.7 mmHg for values less than 100 mmHg, -0.5 +/- 1.8 mmHg for arterial PCO2, and 0.01 +/- 0.01 for pHa. Although the occasional marked discrepancies between one or more CIABG and in vitro values could sometimes be corrected by flushing the arterial catheter or repositioning the sensor, usually we could not determine the cause of the discrepancy or which values were the more accurate.

CONCLUSIONS

Over the range of values and under the clinical conditions studied, CIABG monitoring provides immediate blood gas results and trend information with sufficient agreement with in vitro results to be reliable for decision making in most clinical circumstances. Generally, the differences in the values between the two methods of analysis were the result of the combination of the inherent errors of each method. Additional studies need to be undertaken to evaluate the performance of the CIABG monitor across wider ranges of blood gas values, especially for arterial PO2 values less than 60 mmHg and arterial PCO2 values greater than 50 mmHg.

摘要

背景

我们将一种新型的连续动脉血气(CIABG)监测仪的性能与定期采集并通过传统设备分析的动脉血气值进行了比较。

方法

使用了一种CIABG监测系统,该系统由一个无菌、一次性使用的光纤传感器以及一个带有内置校准单元和可拆卸显示面板的微处理器控制监测仪组成。传感器通过一根20G的桡动脉导管插入。光线从监测仪传输到传感器尖端,在那里它与对氧气或氢离子(分析物)敏感的荧光染料发生反应。每20秒测量一次由分析物引起的光致发光辐射强度变化,并显示导出的血气值。在三个机构(斯坦福大学医院、埃文斯顿纪念医院和帕洛阿尔托退伍军人管理局医院)之一的29例外科或重症监护病房患者中对29个传感器进行了评估。在手术室研究的持续时间平均为6小时(5 - 8小时),在重症监护病房为46小时(7 - 121小时)。总共将552个值与定期采集并在医院血气实验室分析的值进行了比较。确定了平均偏差(实验室值与CIABG之间的平均差异)、精密度(差异的标准差)和漂移(偏差随时间的变化)。

结果

在动脉血氧分压(PO2)值为32 - 528 mmHg时,平均偏差为 - 1%,这意味着CIABG监测仪的平均测量值比通过传统设备获得的值低1%。精密度为15%。在动脉PO₂值为32 - 99 mmHg时,平均偏差和精密度为 - 0.3 ± 8.9 mmHg。在动脉血二氧化碳分压(PCO₂)值为24 - 54 mmHg时,平均偏差和精密度为1.3 ± 3.3 mmHg,在pHa值为7.23 - 7.57时,平均偏差和精密度为0.01 ± 0.04。观察到的动脉PO₂每日漂移为 - 1.2%,动脉PCO₂为0.3 mmHg,pH为0.01。在两对同类体外血气分析仪中比较的样本,动脉PO₂在整个范围内的偏差和精密度为0.4 ± 4.6%,对于小于100 mmHg的值为0.4 ± 3.7 mmHg,动脉PCO₂为 - 0.5 ± 1.8 mmHg,pHa为0.01 ± 0.01。尽管有时通过冲洗动脉导管或重新放置传感器可以纠正一个或多个CIABG值与体外值之间偶尔出现的明显差异,但通常我们无法确定差异的原因或哪个值更准确。

结论

在所研究的值范围和临床条件下,CIABG监测可提供即时的血气结果和趋势信息,与体外结果有足够的一致性,在大多数临床情况下可用于可靠的决策。一般来说,两种分析方法之间的值差异是每种方法固有误差综合作用的结果。需要进行更多研究来评估CIABG监测仪在更广泛的血气值范围内的性能,特别是对于动脉PO₂值小于60 mmHg和动脉PCO₂值大于50 mmHg的情况。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验