Kadakia S C, Cassaday M, Shaffer R T
Gastroenterology Service, Brooke Army Medical Center, San Antonio, Texas.
Am J Gastroenterol. 1992 Nov;87(11):1594-7.
Percutaneous endoscopic gastrostomy provides a nonsurgical alternative to enteral feeding. However, the percutaneous endoscopic gastrostomy tube may deteriorate, malfunction, or be accidentally expelled from the stomach, requiring replacement. This prospective study was performed to evaluate the use of an all silicone Foley catheter (Foley) as a replacement feeding gastrostomy in 28 patients requiring replacement gastrostomy. A plastic ring and a retention disc were always placed over the Foley prior to replacement. Foley functioned well without replacement in 19 (68%) patients for a mean of 167 days. It needed to be replaced in nine (32%) patients due to malfunction after a mean of 138 days. Lack of migration of Foley was the most striking finding of our study, in contrast to case reports in the literature. These data suggest that Foley can be safely used as replacement gastrostomy tube. A randomized controlled trial comparing the Foley catheter as a replacement tube with other commercially available devices is needed.
经皮内镜下胃造口术为肠内营养提供了一种非手术替代方法。然而,经皮内镜下胃造口管可能会变质、出现故障或意外从胃中脱出,需要更换。本前瞻性研究旨在评估在28例需要更换胃造口术的患者中使用全硅胶Foley导管(Foley)作为替代喂养胃造口管的情况。在更换之前,始终会在Foley导管上放置一个塑料环和一个固定盘。19例(68%)患者的Foley导管无需更换,平均使用167天。9例(32%)患者由于导管故障,平均在138天后需要更换。与文献中的病例报告相比,Foley导管未发生移位是我们研究中最显著的发现。这些数据表明,Foley导管可安全用作替代胃造口管。需要进行一项随机对照试验,将Foley导管作为替代管与其他市售装置进行比较。