[NeuFlex Silastic implant in metacarpophalangeal joint arthroplasty].

作者信息

Erdogan A, Weiss A P C

机构信息

Departement of Orthopaedics, Brown Medical School, Rhode Island Hospital, Providence, RI 02905, USA.

出版信息

Orthopade. 2003 Sep;32(9):789-93. doi: 10.1007/s00132-003-0524-0.

Abstract

The NeuFlex joint implant is, like the Swanson prosthesis, a silicone-spacer-implant for metacarpophalangeal arthroplasty in rheumatoid arthritis or osteoarthritis. The implant design differs essentially through a preformed, 30 degrees anatomically neutral angle and palmar lying movement center, which reduces strain across the hinge by 35%. The Swanson prosthesis is straight with a dorsal lying movement center. The advantage of this new design is an improved overall function arc and range of movement in the MCP-joint with a better grip-function. Results of simulator and laboratory tests are presented and interpreted. In 50 patients (168 MCP-joints), early results show an average functional arc of movement of 61 degrees and an average extensor lag of 12 degrees. The 30 degrees pre-bend angle, corresponding to the normal resting angle of the hand, and the improved flexor tendon excursion and moment arm appear to be the main reasons for the substantially improved flexion following use of the NeuFlex device.

摘要

NeuFlex关节植入物与斯旺森假体一样,是一种用于类风湿性关节炎或骨关节炎掌指关节置换术的硅胶间隔植入物。该植入物的设计主要区别在于有一个预成型的、30度解剖学中性角度以及位于手掌侧的运动中心,这可将铰链处的应力降低35%。斯旺森假体是直的,运动中心位于背侧。这种新设计的优点是掌指关节的整体功能弧和活动范围得到改善,抓握功能更好。文中展示并解读了模拟器和实验室测试的结果。在50名患者(168个掌指关节)中,早期结果显示平均功能活动弧为61度,平均伸肌滞后为12度。与手部正常静息角度相对应的30度预弯角度,以及改善的屈肌腱偏移和力臂,似乎是使用NeuFlex装置后屈曲显著改善的主要原因。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验