Amikam S, Peleg H, Lemer J, Riss E
Br Heart J. 1977 Nov;39(11):1279-81. doi: 10.1136/hrt.39.11.1279.
A patient is described in whom myopotentials orginating from the anterior abdominal wall muscle suppressed the implanted demand pacemaker despite its bipolar mode of action. This phenomenon was shown by simultaneous recording of the electrocardiogram the electromyogram. At operation, a defect in the insulation of a previously repaired epicardial electrode was found lying in close proximity to these muscles. After repair of the insulation defect, normal pacemaker function was restored. It is suggested that the myopotentials leaked into the pacing system through the insulation defect, thereby suppressing the demand unit, which maintained its bipolar mode of pacing throughout.