BVI超声角膜测厚仪的可重复性和再现性。

Repeatability and reproducibility of the BVI ultrasonic Pachymeter.

作者信息

Gunvant P, Broadway D C, Watkins R J

机构信息

Department of Optometry and Ophthalmic Dispensing Anglia Polytechnic University Cambridge, UK.

出版信息

Eye (Lond). 2003 Oct;17(7):825-8. doi: 10.1038/sj.eye.6700485.

Abstract

PURPOSE

To evaluate the interobserver and intraobserver reliability of a commercially available, portable, ultrasonic pachymeter.

METHODS

For the interobserver variability study, 42 healthy subjects underwent repeated ultrasonic pachymetry under topical anaesthesia performed by two observers. For the intraobserver study, another 30 further healthy subjects underwent repeated pachymetry by one of the observers. Agreement was analysed by means of Bland-Altman plots and by determination of the intraclass correlation coefficient.

RESULTS

For the interobserver variability study, the mean measurement difference between observers was 0.7 microm (95% CI -0.8-2.2 microm) and the intraclass correlation coefficient was 0.9958 (95% CI 0.9922-0.9977). The Bland-Altman plot showed narrow limits of agreement with respect to central corneal thickness (CCT). For the intraobserver variability study, the mean difference between the repeated measurements was 0.9 microm (95% CI -3.1-1.3 microm). The intraclass correlation coefficient was 0.9934 (95% CI 0.9863-0.9969). A Bland-Altman plot again showed narrow limits of agreement with respect to CCT. The mean CCT for 72 subjects was 538 microm (95% CI 528-545 microm).

CONCLUSIONS

Measurements of CCT using the BVI Pocket Pachymeter were repeatable and had excellent interobserver reliability. Measurement variation amounted to less than 0.2% assuming a mean CCT of 538 microm.

摘要

目的

评估一种市售便携式超声角膜测厚仪的观察者间和观察者内可靠性。

方法

在观察者间变异性研究中,42名健康受试者在两名观察者进行的表面麻醉下接受重复超声角膜测厚。在观察者内研究中,另外30名健康受试者由一名观察者进行重复角膜测厚。通过Bland-Altman图和组内相关系数的测定来分析一致性。

结果

在观察者间变异性研究中,观察者之间的平均测量差异为0.7微米(95%CI -0.8 - 2.2微米),组内相关系数为0.9958(95%CI 0.9922 - 0.9977)。Bland-Altman图显示中央角膜厚度(CCT)的一致性界限较窄。在观察者内变异性研究中,重复测量之间的平均差异为0.9微米(95%CI -3.1 - 1.3微米)。组内相关系数为0.9934(95%CI 0.9863 - 0.9969)。Bland-Altman图再次显示CCT的一致性界限较窄。72名受试者的平均CCT为538微米(95%CI 528 - 545微米)。

结论

使用BVI Pocket角膜测厚仪测量CCT具有可重复性且观察者间可靠性极佳。假设平均CCT为538微米,测量变化小于0.2%。