在评估中央角膜厚度的超声测厚测量中观察者内和观察者间的可重复性。

Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness.

作者信息

Miglior S, Albe E, Guareschi M, Mandelli G, Gomarasca S, Orzalesi N

机构信息

Department of Ophthalmology, University of Milan Bicocca, Department of Neurosciences and Biomedical Technologies, Via Cadore 48, 20052 Monza, Milan, Italy.

出版信息

Br J Ophthalmol. 2004 Feb;88(2):174-7. doi: 10.1136/bjo.2003.023416.

Abstract

PURPOSE

To assess reproducibility of central corneal thickness (CCT) measurement by means of ultrasonic pachymetry.

METHODS

Fifty one volunteers underwent three sessions of CCT measurements, each consisting of three CCT measurements, performed by each of three different observers. Intra- and interobserver reproducibility was calculated by means of intraclass correlation coefficient (ICC). The expected range of variability between two independent evaluations was calculated using scatter plots of each test-retest difference against their mean. The standard deviation of the mean differences in the test-retest scores was used to describe the differences in the score spread.

RESULTS

The ICC ranges of the intra- and interobserver evaluations were 0.95-0.97 and 0.89-0.95 respectively; the expected variability was < or = +/-1% and < or = +/- 2% respectively (95% confidence interval).

CONCLUSIONS

The measurement of CCT by means of ultrasonic pachymetry is highly reproducible.

摘要

目的

通过超声测厚法评估中央角膜厚度(CCT)测量的可重复性。

方法

51名志愿者接受了三次CCT测量,每次测量包括由三名不同观察者分别进行的三次CCT测量。通过组内相关系数(ICC)计算观察者内和观察者间的可重复性。使用每次重测差异与其平均值的散点图计算两次独立评估之间的预期变异范围。重测分数平均差异的标准差用于描述分数分布的差异。

结果

观察者内和观察者间评估的ICC范围分别为0.95 - 0.97和0.89 - 0.95;预期变异分别≤±1%和≤±2%(95%置信区间)。

结论

通过超声测厚法测量CCT具有高度可重复性。

相似文献

7
Central corneal thickness of Korean patients with glaucoma.韩国青光眼患者的中央角膜厚度
Ophthalmology. 2007 May;114(5):927-30. doi: 10.1016/j.ophtha.2006.09.036. Epub 2007 Feb 21.

引用本文的文献

8
Effect of Demographic Variables on the Regional Corneal Pachymetry.人口统计学变量对区域角膜厚度的影响。
Asia Pac J Ophthalmol (Phila). 2019 Jul-Aug;8(4):324-329. doi: 10.1097/APO.0000000000000252.
9
Interdevice variability of central corneal thickness measurement.不同仪器测量中央角膜厚度的变异性。
PLoS One. 2018 Sep 13;13(9):e0203884. doi: 10.1371/journal.pone.0203884. eCollection 2018.

本文引用的文献

2
[Applanation tonometry].[压平眼压测量法]
Ophthalmologica. 1957 Oct;134(4):221-42. doi: 10.1159/000303213.
10
Increased corneal thickness in patients with ocular hypertension.高眼压症患者角膜厚度增加。
Arch Ophthalmol. 2001 Mar;119(3):334-6. doi: 10.1001/archopht.119.3.334.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验