Byun Seok Soo, Kim Hyeon Hoe, Lee Eunsik, Paick Jae-Seung, Kamg Weechang, Oh Seung-June
Department of Urology, Seoul National University College of Medicine, Seoul, South Korea.
Urology. 2003 Oct;62(4):656-60. doi: 10.1016/s0090-4295(03)00559-4.
To investigate the accuracy of a portable three-dimensional hand-held BladderScan (BS) and two-dimensional conventional ultrasonography (CUS) in the estimation of bladder volumes.
The bladder volume of 65 subjects was measured during filling cystometry using two ultrasound modalities. Serial measurements were performed when the infused volume reached 100, 200, 300, and 400 mL; the investigator measured the volumes three times each with BS and CUS. Each corresponding true volume was calculated by interpolation using zero volume and the final catheterized volume at the end of cystometry. The accuracy of the two ultrasound methods was compared using raw scaled values of the volumes and the percentage of differences of volume.
Bladder volumes determined by CUS and BS were underestimated by 21.8% and 3.3% (mean values), respectively. The percentage of differences of volume of BS was not statistically significant (P >0.05), but that of CUS was statistically significant (P <0.00001). No significant percentage of differences of volume change according to the volume status was observed by either ultrasound measurement method (P >0.05). Linear regression analysis between the true volumes and the raw scaled volumes showed that the slope of the BS was closer to 1.0 than that of the CUS, and the difference was statistically significant (P <0.001). The effects of diagnosis, age, sex, and body mass index were not significant in determining the bladder volume by either method (P >0.05).
Our results demonstrated that a three-dimensional hand-held scanner measures the bladder volume in a reproducible and accurate manner for a wide range of bladder volumes and is superior to two-dimensional stationary ultrasonography.
研究便携式三维手持膀胱扫描仪(BS)和二维传统超声检查(CUS)在估计膀胱容量方面的准确性。
在膀胱充盈测压期间,使用两种超声检查方式测量65名受试者的膀胱容量。当注入量达到100、200、300和400 mL时进行系列测量;研究者分别用BS和CUS对容量测量三次。每个相应的真实容量通过使用零容量和膀胱测压结束时最终导尿的容量进行插值计算得出。使用容量的原始标度值和容量差异百分比比较两种超声检查方法的准确性。
CUS和BS测定的膀胱容量分别平均低估了21.8%和3.3%。BS的容量差异百分比无统计学意义(P>0.05),但CUS的容量差异百分比有统计学意义(P<0.00001)。两种超声测量方法均未观察到根据容量状态的容量变化差异百分比有显著变化(P>0.05)。真实容量与原始标度容量之间的线性回归分析表明,BS的斜率比CUS更接近1.0,且差异有统计学意义(P<0.001)。诊断、年龄、性别和体重指数对两种方法测定膀胱容量的影响均不显著(P>0.05)。
我们的结果表明,三维手持扫描仪能够以可重复且准确的方式测量广泛膀胱容量范围内的膀胱容量,并且优于二维静态超声检查。