百合 α-200 便携式超声膀胱扫描仪测量膀胱容量的准确性评估。

Feasibility of the Lilium α-200 portable ultrasound bladder scanner for accurate bladder volume measurement.

机构信息

Department of Urology, Nagoya University Graduate School of Medicine, Nagoya, Japan.

Department of Urology, Oita Oka Hospital, Keiwakai, Japan.

出版信息

Investig Clin Urol. 2020 Nov;61(6):613-618. doi: 10.4111/icu.20200012.

Abstract

PURPOSE

The aim of this study was to investigate whether data obtained from the Lilium α-200 (Lilium Otsuka Co., Ltd., Japan) correlated with conventional frequency-volume chart (FVC) and post-void residual urine volume (PVR) obtained by urethral catheterization.

MATERIALS AND METHODS

This was a prospective multicentre study. Patients hospitalized for the treatment of lower urinary tract symptoms were included. Patients were evaluated with conventional FVC and Lilium α-200 for 2 days. PVR was measured by urethral catherization after urination at the end of the 2 day evaluation period.

RESULTS

A total of 42 patients were enrolled in this study. Voided volume and PVR measured by Lilium α-200 were significantly correlated with voided volume obtained from conventional FVC and PVR measured by urethral catheterization, respectively. There was considerable measurement error in voided volume measured by Lilium α-200 (-21.0±102.0 mL). In contrast, the error between PVR measured by the Lilium α-200 and PVR obtained by urethral catheterization was 2.4±52.0 mL. Additionally, high body mass index, but not sex, benign prostate hyperplasia, time zone of measurement (daytime vs. nighttime), and examiners (a urologist versus other healthcare providers) were significantly associated with inaccurate results in voided volume.

CONCLUSIONS

Voided volume and PVR measured by the Lilium α-200 were correlated with voided volume obtained from conventional FVC and PVR measured by urethral catheterization, although accuracy of the measurements was not high. The Lilium α-200 is a useful device to easily measure approximate bladder volume.

摘要

目的

本研究旨在探讨 Lilium α-200(日本大塚制药株式会社)获得的数据是否与通过尿道导管获得的传统频率体积图(FVC)和排尿后残余尿量(PVR)相关。

材料和方法

这是一项前瞻性多中心研究。纳入因下尿路症状住院治疗的患者。患者在 2 天内接受传统 FVC 和 Lilium α-200 评估。在 2 天评估期结束时排尿后,通过尿道导管测量 PVR。

结果

本研究共纳入 42 例患者。Lilium α-200 测量的尿量和 PVR 与传统 FVC 获得的尿量和尿道导管测量的 PVR 均呈显著相关。Lilium α-200 测量的尿量存在较大的测量误差(-21.0±102.0 mL)。相比之下,Lilium α-200 测量的 PVR 与尿道导管测量的 PVR 之间的误差为 2.4±52.0 mL。此外,高体重指数而非性别、良性前列腺增生、测量时间(白天与夜间)和检查者(泌尿科医生与其他医疗保健提供者)与尿量测量结果不准确显著相关。

结论

Lilium α-200 测量的尿量和 PVR 与传统 FVC 获得的尿量和尿道导管测量的 PVR 相关,但测量的准确性不高。Lilium α-200 是一种方便测量近似膀胱容量的有用设备。