Accuracy and precision of a new portable ultrasound scanner, the BME-150A, in residual urine volume measurement: a comparison with the BladderScan BVI 3000.

作者信息

Choe Jin Ho, Lee Ji Yeon, Lee Kyu-Sung

机构信息

Department of Urology, Cheil General Hospital, Sungkyunkwan University School of Medicine, Seoul, South Korea.

出版信息

Int Urogynecol J Pelvic Floor Dysfunct. 2007 Jun;18(6):641-4. doi: 10.1007/s00192-006-0219-2. Epub 2006 Nov 7.

Abstract

The objective of the study was to determine the relative accuracy of a new portable ultrasound unit, BME-150A, and the BladderScan BVI 3000, as assessed in comparison with the catheterized residual urine volume. We used both of these machines to prospectively measure the residual urine volumes of 89 patients (40 men and 49 women) who were undergoing urodynamic studies. The ultrasound measurements were compared with the post-scan bladder volumes obtained by catheterization in the same patients. The ultrasounds were followed immediately (within 5 min) by in-and-out catheterizations while the patients were in a supine position. There were a total of 116 paired measurements made. The BME-150A and the BVI 3000 demonstrated a correlation with the residual volume of 0.92 and 0.94, and a mean difference from the true residual volume of 7.8 and 3.6 ml, respectively. Intraclass correlation coefficients for the accuracy of the two bladder scans were 0.90 for BME-150A and 0.95 for BVI 3000. The difference of accuracy between the two models was not significant (p = 0.2421). There were six cases in which a follow-up evaluation of falsely elevated post-void residual urine volume measurements on the ultrasound studies resulted in comparatively low catheterized volumes, with a range of differences from 66 to 275.5 ml. These cases were diagnosed with an ovarian cyst, uterine myoma, or uterine adenomyosis on pelvic ultrasonography. The accuracy of the BME-150A is comparable to that of the BVI 3000 in estimating the true residual urine volumes and is sufficient enough for us to recommend its use as an alternative to catheterization.

摘要

本研究的目的是确定一款新型便携式超声设备BME - 150A与膀胱扫描仪BladderScan BVI 3000相对于导尿测定残余尿量的相对准确性。我们使用这两款设备对89例正在接受尿动力学检查的患者(40名男性和49名女性)的残余尿量进行前瞻性测量。将超声测量结果与同一患者导尿后获得的膀胱容积进行比较。患者仰卧位时,超声检查后立即(5分钟内)进行进出式导尿。总共进行了116次配对测量。BME - 150A和BVI 3000与残余尿量的相关性分别为0.92和0.94,与真实残余尿量的平均差异分别为7.8毫升和3.6毫升。两种膀胱扫描准确性的组内相关系数,BME - 150A为0.90,BVI 3000为0.95。两种型号之间的准确性差异不显著(p = 0.2421)。有6例患者在超声检查中出现排尿后残余尿量测量值假性升高的后续评估结果,导致导尿测量值相对较低,差异范围为66至275.5毫升。盆腔超声检查诊断这些病例患有卵巢囊肿、子宫肌瘤或子宫腺肌病。在估计真实残余尿量方面,BME - 150A的准确性与BVI 3000相当,足以推荐其作为导尿的替代方法使用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验