高度交联聚乙烯在全髋关节置换术中两年的临床性能:一项随机前瞻性试验。

Clinical performance of a highly cross-linked polyethylene at two years in total hip arthroplasty: a randomized prospective trial.

作者信息

Martell John M, Verner James J, Incavo Stephen J

机构信息

Department of Orthopaedics, The University of Chicago, Illinois 60637, USA.

出版信息

J Arthroplasty. 2003 Oct;18(7 Suppl 1):55-9. doi: 10.1016/s0883-5403(03)00341-3.

Abstract

We report the 2-year results for a prospective randomized trial comparing highly cross-linked with standard polyethylene in total hip replacements. In our study, 46 hips were available for radiographic analysis at 2- and 3-year follow-up. Femoral bearings were 28-mm cobalt chrome with the polyethylene insert randomly selected at the time of implantation to be highly cross-linked polyethylene (Crossfire; Stryker Howmedica Osteonics, Allendale, NJ) or standard polyethylene (N(2)/Vac, Stryker Howmedica Osteonics). Polyethylene wear rates were measured based on anterior-posterior (AP) and lateral pelvis radiographs at 6 weeks and at yearly intervals using a validated computer-assisted edge-detection method. Wear rates between the 2 groups were compared using the nonparametric Mann-Whitney test at the 95% level. A significant reduction in 2- and 3-dimensional linear wear rates (42% and 50%) was found in the highly cross-linked group (P =.001 and P =.005).

摘要

我们报告了一项前瞻性随机试验的2年结果,该试验比较了全髋关节置换术中高交联聚乙烯与标准聚乙烯的效果。在我们的研究中,46个髋关节在2年和3年随访时可进行影像学分析。股骨假体为28毫米钴铬合金,聚乙烯内衬在植入时随机选择,分别为高交联聚乙烯(Crossfire;史赛克豪美医疗,新泽西州阿伦代尔)或标准聚乙烯(N(2)/Vac,史赛克豪美医疗)。使用经过验证的计算机辅助边缘检测方法,根据前后位(AP)和骨盆侧位X线片在6周及每年测量聚乙烯磨损率。两组之间的磨损率采用非参数曼-惠特尼检验在95%水平进行比较。高交联组在二维和三维线性磨损率方面均有显著降低(分别为42%和50%)(P = 0.001和P = 0.005)。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验