[ABL 510上同时测量pO2和sO2的相关性]

[Pertinence of simultaneous measurements of pO2 and sO2 on ABL 510].

作者信息

Gouget B, Manene D, Andrimahatratra R, Bogard M, Gourmelin Y

机构信息

CNEH, Paris, France.

出版信息

Ann Biol Clin (Paris). 1992;50(4):247-50.

PMID:1456503
Abstract

Blood tonometry is the only method to assess the accuracy of the pO2 determination on blood gas analyzers. ABL instruments by Radiometer were tested by two types of tonometry (film and bubble tonometry) and the validity of the algorithm for pO2 correction was analysed with these results. The role of the presentation of the specimen is also discussed. For the precision study on the ABL 510 analyzer, coefficients of variation for pO2 were < 0.37% and < 1.7% for within-run and day-to-day series respectively. pO2 accuracy was excellent. Linearity was verified between 0-620 mmHg (82.5 kPa), and interinstrument comparisons demonstrated a strict correlation with the Ciba-Corning 178 instrument. ABL 510 measures simultaneously oxygen saturation by spectrophotometry. Analytical results showed an acceptable level of imprecision, but the definition and the clinical significance of this parameter are ambiguous.

摘要

血液张力测定法是评估血气分析仪上氧分压(pO2)测定准确性的唯一方法。Radiometer公司的ABL仪器通过两种类型的张力测定法(薄膜和气泡张力测定法)进行了测试,并根据这些结果分析了pO2校正算法的有效性。还讨论了样本呈现方式的作用。对于ABL 510分析仪的精密度研究,批内和日间系列的pO2变异系数分别<0.37%和<1.7%。pO2准确性极佳。在0 - 620 mmHg(82.5 kPa)之间验证了线性,仪器间比较表明与Ciba - Corning 178仪器有严格的相关性。ABL 510通过分光光度法同时测量氧饱和度。分析结果显示不精密度处于可接受水平,但该参数的定义和临床意义尚不明确。