Warkentin B, Steciw S, Rathee S, Fallone B G
Department of Medical Physics, Cross Cancer Institute, University of Alberta, 11560 University Avenue, Edmonton, Alberta T6G 1Z2, Canada.
Med Phys. 2003 Dec;30(12):3143-55. doi: 10.1118/1.1625440.
A convolution-based calibration procedure has been developed to use an amorphous silicon flat-panel electronic portal imaging device (EPID) for accurate dosimetric verification of intensity-modulated radiotherapy (IMRT) treatments. Raw EPID images were deconvolved to accurate, high-resolution 2-D distributions of primary fluence using a scatter kernel composed of two elements: a Monte Carlo generated kernel describing dose deposition in the EPID phosphor, and an empirically derived kernel describing optical photon spreading. Relative fluence profiles measured with the EPID are in very good agreement with those measured with a diamond detector, and exhibit excellent spatial resolution required for IMRT verification. For dosimetric verification, the EPID-measured primary fluences are convolved with a Monte Carlo kernel describing dose deposition in a solid water phantom, and cross-calibrated with ion chamber measurements. Dose distributions measured using the EPID agree to within 2.1% with those measured with film for open fields of 2 x 2 cm2 and 10 x 10 cm2. Predictions of the EPID phantom scattering factors (SPE) based on our scatter kernels are within 1% of the SPE measured for open field sizes of up to 16 x 16 cm2. Pretreatment verifications of step-and-shoot IMRT treatments using the EPID are in good agreement with those performed with film, with a mean percent difference of 0.2 +/- 1.0% for three IMRT treatments (24 fields).
已开发出一种基于卷积的校准程序,以使用非晶硅平板电子射野影像装置(EPID)对调强放射治疗(IMRT)进行精确的剂量验证。利用由两个元素组成的散射核,将原始EPID图像进行去卷积处理,以获得精确的高分辨率二维原射线注量分布:一个由蒙特卡罗生成的描述EPID磷光体中剂量沉积的核,以及一个根据经验得出的描述光学光子扩散的核。用EPID测量的相对注量剖面与用金刚石探测器测量的结果非常吻合,并且具有IMRT验证所需的出色空间分辨率。为了进行剂量验证,将EPID测量的原射线注量与一个描述固体水模体中剂量沉积的蒙特卡罗核进行卷积,并与电离室测量结果进行交叉校准。对于2×2 cm2和10×10 cm2的开放射野,用EPID测量的剂量分布与用胶片测量的结果在2.1%以内相符。基于我们的散射核对EPID模体散射因子(SPE)的预测,对于高达16×16 cm2的开放射野尺寸,其结果在测量的SPE的1%以内。使用EPID对静态调强放射治疗进行治疗前验证,与用胶片进行的验证结果良好相符,对于三次IMRT治疗(24个射野),平均百分比差异为0.2±1.0%。