Kullenberg Ragnar
Department of Radiology, County Hospital, SE-301 85 Halmstad, Sweden.
J Clin Densitom. 2003 Winter;6(4):367-72. doi: 10.1385/jcd:6:4:367.
The new dual X-ray and laser technology (DXL) gives a more accurate determination of bone mineral density (BMD) than ordinary dual X-ray absorptiometry (DXA) technology because of the ability to eliminate fat tissue both inside and outside the measured bone. In this study the reference database for BMD measured at the calcaneus by the DXL Calscan device is reported. The database was obtained from 993 healthy women and 459 healthy men in a population from southern Sweden. Inclusion criteria were: healthy Swedish Caucasians, 15-85 yr of age for women and 19-85 yr of age for men, no history of osteoporosis treatment, no use of corticosteroids for more than 3 mo, and no extended bed rest. The young adult reference mean BMD for women was found to be 0.483 +/- 0.062 g/cm2 and for men 0.556 +/- 0.074 g/cm2. The age-adjusted odds ratio was 3.7 for a history of fracture among women aged 50 yr and over, comparing subjects with a 1-SD reduction in bone density to subjects with a bone density above this value. The DXL Calscan device used for the study was calibrated weekly against a heel bone phantom. The precision of these measurements was 0.5%. The in vivo precision was 1.2%, as assessed by duplicate measurements on 35 healthy individuals (mean age 52 yr, range 25-72 yr).
新型双能X射线与激光技术(DXL)能够消除被测骨骼内外的脂肪组织,因此与普通双能X线吸收法(DXA)技术相比,能更准确地测定骨矿物质密度(BMD)。本研究报告了通过DXL Calscan设备测量跟骨BMD的参考数据库。该数据库来自瑞典南部人群中的993名健康女性和459名健康男性。纳入标准为:健康的瑞典高加索人,女性年龄在15至85岁之间,男性年龄在19至85岁之间,无骨质疏松治疗史,未使用皮质类固醇超过3个月,且无长期卧床休息。女性青年成人参考平均BMD为0.483±0.062g/cm²,男性为0.556±0.074g/cm²。在50岁及以上的女性中,将骨密度降低1个标准差的受试者与骨密度高于此值的受试者进行比较,骨折史的年龄调整比值比为3.7。用于该研究的DXL Calscan设备每周根据足跟骨模型进行校准。这些测量的精密度为0.5%。通过对35名健康个体(平均年龄52岁,范围25至72岁)进行重复测量评估,体内精密度为1.2%。