Suppr超能文献

Accuracy, reproducibility and clinical utility of the CoaguChek S portable international normalized ratio monitor in an outpatient anticoagulation clinic.

作者信息

Jackson S L, Bereznicki L R, Peterson G M, Marsden K A, Jupe D M L, Tegg E, Vial J H, Kimber R I

机构信息

School of Pharmacy, University of Tasmania, Tasmania, Australia.

出版信息

Clin Lab Haematol. 2004 Feb;26(1):49-55. doi: 10.1111/j.0141-9854.2003.00578.x.

Abstract

The accuracy and reproducibility of the CoaguChek S, and its clinical agreement with conventional laboratory international normalized ratio (INR) determination, were evaluated in an outpatient anticoagulation clinic setting. Forty-three patients provided 248 paired INR measurements for analysis. The paired results were highly correlated (r = 0.90). The mean coefficient of variation for the CoaguChek S for a random sample of 21 patients with three repeated tests each, was 4%. Clinical applicability was also measured by discrepant INR values, as defined in the literature by expanded and narrow agreement, and by INR values resulting in a different clinical decision by a blinded haematology registrar. Expanded agreement and narrow agreement between the two INR values occurred 90 and 88% of the time, respectively. The stricter criteria set down by the clinician resulted in 73% of paired results producing the same dosage decision. The CoaguChek S displayed good correlation with laboratory determination of INR and compared relatively well with expanded and narrow clinical agreement criteria.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验