使用A超超声检查法测量儿童眼部成分的变异性。

Variability of the ocular component measurements in children using A-scan ultrasonography.

作者信息

Kurtz Daniel, Manny Ruth, Hussein Mohamed

机构信息

New England College of Optometry, Boston, Massachusetts 02115, USA.

出版信息

Optom Vis Sci. 2004 Jan;81(1):35-43. doi: 10.1097/00006324-200401000-00008.

Abstract

Knowledge of the precision with which the various ocular components can be measured with available techniques is vital to our ability to track changes in the anatomy of the eye in relation to the development of refractive error. Corneal touch A-scan ultrasound measurements were obtained by either an instrument-mounted or by a hand-held method on 469 children aged 6 to 11 years at their enrollment into the Correction of Myopia Evaluation Trial. Variability of measurements was calculated for overall axial length, anterior chamber depth, lens thickness, and vitreous chamber depth. The mean variability of overall axial length was 0.062 +/- 0.043 mm and 0.061 +/- 0.056 mm for the right and left eyes, respectively. Mean variability of the ocular component measures was similar. The mean variability was not influenced by the age, gender, level of myopia, or residual accommodation of the subjects. Statistically significant differences between the instrument-mounted and the hand-held method and among certain ethnic groups were found, but the differences were not of a magnitude to be clinically significant. A-scan ultrasonography is sensitive to changes in the axial length and vitreous chamber depth equivalent to 0.25 D and is therefore a useful technique to assess changes in these ocular components in children. The precision of lens thickness is poorer than the equivalent of 1.0 D, and, therefore, A-scan may not be sufficiently precise to be useful in studies of active accommodation or lens growth.

摘要

了解使用现有技术测量各种眼部组件的精确程度,对于我们追踪眼睛解剖结构相对于屈光不正发展的变化的能力至关重要。在469名6至11岁儿童纳入近视矫正评估试验时,通过仪器安装或手持方法获得角膜接触A扫描超声测量值。计算了总眼轴长度、前房深度、晶状体厚度和玻璃体腔深度测量值的变异性。右眼和左眼总眼轴长度的平均变异性分别为0.062±0.043毫米和0.061±0.056毫米。眼部组件测量值的平均变异性相似。平均变异性不受受试者的年龄、性别、近视程度或残余调节的影响。发现仪器安装方法和手持方法之间以及某些种族群体之间存在统计学上的显著差异,但差异程度不具有临床意义。A扫描超声检查对眼轴长度和玻璃体腔深度相当于0.25 D的变化敏感,因此是评估儿童这些眼部组件变化的有用技术。晶状体厚度的精确度低于相当于1.0 D的变化,因此,A扫描在研究主动调节或晶状体生长时可能不够精确而无法使用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验