The use of a checklist for anaesthetic machines.

作者信息

Barthram C, McClymont W

机构信息

Department of Anaesthesia, Ninewells Hospital, Dundee.

出版信息

Anaesthesia. 1992 Dec;47(12):1066-9. doi: 10.1111/j.1365-2044.1992.tb04206.x.

Abstract

The use of the Association of Anaesthetists of Great Britain and Ireland checklist for anaesthetic machines, based on an oxygen analyser, was surveyed over a 5-week period in a teaching hospital. Fifty-five completed checklists were analysed; no problems developed during anaesthesia which were missed by the checklist. The mean time taken to complete the checklist for one machine was 8.9 min with a range of 5 to 19 min; for two consecutive machines it was 18.25 min with a range of 10 to 30 min. The most frequent faults detected were the poor reliability of some oxygen analysers, absent ventilator disconnection alarms, and absent oxygen supply failure alarms on some older machines. Faults were found in 60% of the machines checked; 18% of these were deemed to be serious.

摘要

在一家教学医院,对基于氧气分析仪的英国和爱尔兰麻醉医师协会麻醉机检查表的使用情况进行了为期5周的调查。分析了55份完整的检查表;麻醉过程中未出现检查表遗漏的问题。检查一台麻醉机完成检查表的平均时间为8.9分钟,范围为5至19分钟;连续检查两台麻醉机的平均时间为18.25分钟,范围为10至30分钟。检测到的最常见故障是一些氧气分析仪可靠性差、部分呼吸机断开连接警报缺失以及一些旧机器上氧气供应故障警报缺失。在检查的机器中,60%发现了故障;其中18%被认为是严重故障。