Effect of digital cellular phones on tachyarrhythmia analysis of automated external defibrillators.

作者信息

Kanz Karl-Georg, Kay Mike V, Biberthaler Peter, Russ Walter, Lackner Christian K, Mutschler Wolf

机构信息

Institute for Emergency Service and Management in Medicine INM, Munich, Germany.

出版信息

Eur J Emerg Med. 2004 Apr;11(2):75-80. doi: 10.1097/00063110-200404000-00004.

DOI:10.1097/00063110-200404000-00004
PMID:15028895
Abstract

OBJECTIVES

Emergency services personnel, family members, laypersons or patients often carry and use mobile phones on sites of emergencies. As there are reported effects on implanted pacemakers and cardioverter defibrillators, the influence of digital cellular phones on automated external defibrillators was studied.

METHODS

Twelve automated external defibrillator models were bench tested for their correct decision to or not to advise a shock, while being exposed to electromagnetic interference from a handheld cellular phone with 2 W or a portable cellular phone with 8 W transmitting power. The phones were programmed by a special subscriber identity module card to maximum output power with a carrier frequency of 906.2 MHz. The tests were conducted with a burst frequency of 217 Hz in speech mode and 2-8 Hz in discontinuous transmitting exchange mode. The sensitivity and specificity of electrocardiogram analysis systems were tested, with shockable and non-shockable rhythms provided by an electrocardiogram simulator and on two human subjects with normal sinus rhythm.

RESULTS

A total of 8640 tests were recorded, each automated external defibrillator was tested a total of 720 times. The automated external defibrillators demonstrated a sensitivity of 100% and a specificity of 100%, representing a positive likelihood ratio of 8641 and a negative likelihood ratio of 0.000. In this setting all automated external defibrillators analysed correctly even under worst-case testing conditions, and performed excellently without any single failure. In some devices, voice prompts were distorted beyond comprehension, as the coil of the automated external defibrillator speaker received the pulsed signals.

CONCLUSION

Shock advisory systems of automated external defibrillators are not susceptible to electromagnetic interference of 900 MHz cellular phones. Voice prompts, however, could be distorted by the operation of nearby digital mobile phones. During automated external defibrillator training this issue needs to be addressed.

摘要

相似文献

1
Effect of digital cellular phones on tachyarrhythmia analysis of automated external defibrillators.
Eur J Emerg Med. 2004 Apr;11(2):75-80. doi: 10.1097/00063110-200404000-00004.
2
Clinical testing of cellular phone ringing interference with automated external defibrillators.手机铃声对自动体外除颤器干扰的临床测试。
Resuscitation. 2006 Dec;71(3):391-4. doi: 10.1016/j.resuscitation.2006.04.013. Epub 2006 Sep 20.
3
Susceptibility of automated external defibrillators to train overhead lines and metro third rails.自动体外除颤器受架空电线和地铁第三轨影响的敏感性。
Resuscitation. 2004 Aug;62(2):189-98. doi: 10.1016/j.resuscitation.2004.02.018.
4
[Electromagnetic interference between automatic defibrillators and digital and analog cellular telephones].
Rev Esp Cardiol. 1998 May;51(5):375-82. doi: 10.1016/s0300-8932(98)74761-x.
5
Influence of D-net (European GSM-Standard) cellular phones on pacemaker function in 50 patients with permanent pacemakers.D网(欧洲GSM标准)手机对50例永久性起搏器患者起搏器功能的影响。
Pacing Clin Electrophysiol. 1996 Oct;19(10):1456-8. doi: 10.1111/j.1540-8159.1996.tb03158.x.
6
Electromagnetic interference of pacemakers by mobile phones.手机对起搏器的电磁干扰。
Pacing Clin Electrophysiol. 1996 Oct;19(10):1431-46. doi: 10.1111/j.1540-8159.1996.tb03155.x.
7
Electromagnetic interference of external pacemakers by walkie-talkies and digital cellular phones: experimental study.对讲机和数字蜂窝电话对体外起搏器的电磁干扰:实验研究
Pacing Clin Electrophysiol. 1999 Apr;22(4 Pt 1):588-93. doi: 10.1111/j.1540-8159.1999.tb00499.x.
8
Effects of mobile telephones on the function of implantable cardioverter defibrillators.移动电话对植入式心脏复律除颤器功能的影响。
Ann Noninvasive Electrocardiol. 2005 Oct;10(4):409-13. doi: 10.1111/j.1542-474X.2005.00057.x.
9
Reliability of electromagnetic filters of cardiac pacemakers tested by cellular telephone ringing.通过手机铃声测试心脏起搏器电磁滤波器的可靠性。
Heart Rhythm. 2005 Aug;2(8):837-41. doi: 10.1016/j.hrthm.2005.03.011.
10
Mobile phones in the hospital--past, present and future.医院中的手机——过去、现在与未来。