Lau Carolin, Reiter Sabine, Schuhmann Wolfgang, Gründler Peter
Fachbereich Chemie, Abteilung für Analytische, Technische und Umweltchemie, Universität Rostock, Albert-Einstein-Strasse 3a, 18051, Rostock, Germany.
Anal Bioanal Chem. 2004 May;379(2):255-60. doi: 10.1007/s00216-004-2585-2. Epub 2004 Mar 23.
Heated electrodes were applied for the non-isothermal operation of amperometric glucose biosensors based on glucose oxidase immobilised on the electrode surface by entrapment within a polymer layer. The localised deposition of the polymer film under simultaneous entrapment of the enzyme was achieved by an electrochemically induced pH-modulation in the diffusion zone in front of the electrode, thus altering the solubility of the polymer chains. This non-manual sensor preparation protocol could be successfully used for the modification of a novel indirectly heated electrode. The non-isothermal operating mode allows working at the optimum temperature of the enzyme sensors without any thermal distortion of the bulk solution. Increased surface temperature of the sensor thus accelerates transport as well as kinetic processes, resulting in an enhanced amperometric signal. In the presence of interfering compounds such as ascorbic acid, the proposed technique allows use of the diverging thermal impact on the sensing process, for different electrochemically active compounds, for a deconvolution of the amperometric signal at different electrode temperatures. A calculation method for determination of glucose in the presence of one interfering compound is presented as a basis for a calculative interference elimination.