Sacu Stefan, Menapace Rupert, Findl Oliver, Georgopoulos Michael, Buehl Wolf, Kriechbaum Katharina, Rainer Georg
Department of Ophthalmology, University of Vienna, Vienna, Austria.
J Cataract Refract Surg. 2004 Mar;30(3):658-62. doi: 10.1016/j.jcrs.2003.07.005.
To evaluate the role of posterior optic edge design and the effect of anterior capsule polishing on peripheral fibrotic posterior capsule opacification (PCO).
Department of Ophthalmology, University of Vienna, Vienna, Austria.
This randomized prospective study comprised 144 eyes of 72 patients with bilateral age-related cataract. Each patient had standardized cataract surgery in both eyes by the same surgeon. Group 1 (46 patients) received a round-edged hydrophobic acrylic IOL (AMO Sensar AR40) in 1 eye and a sharp-edged hydrophobic acrylic IOL (AMO Sensar OptiEdge AR40e) in the other eye. Group 2 (26 patients) received a silicone IOL (Pharmacia CeeOn 911A) with a truncated optic in both eyes. In this group, the anterior capsule was extensively polished in 1 eye and was left unpolished in the other eye. Digital slitlamp photographs were taken 1 year postoperatively using a standardized photographic technique for fibrotic PCO. The intensity of PCO was subjectively graded (score 0 to 4) by 2 masked examiners.
Subjective PCO scores correlated well between the 2 examiners (r = 0.88). In Group 1, the mean PCO score was 0.26 for the OptiEdge AR40e IOL and 0.90 for the AR40 IOL (P<.01). In Group 2, the mean PCO score was 0.24 in eyes with a polished capsule and 0.17 in eyes in which the capsule was not polished (P =.31).
The sharp-edged OptiEdge AR40e IOL led to significantly less peripheral fibrotic PCO 1 year postoperatively than the round-edged AR40 IOL. In eyes with the sharp-edged silicone 911A IOL, anterior capsule polishing caused no significant difference in fibrotic PCO.
评估后光学边缘设计的作用以及前囊抛光对周边纤维化后囊混浊(PCO)的影响。
奥地利维也纳大学眼科。
这项随机前瞻性研究纳入了72例双侧年龄相关性白内障患者的144只眼。每位患者的双眼均由同一位外科医生进行标准化白内障手术。第1组(46例患者)一只眼植入圆边疏水丙烯酸人工晶状体(AMO Sensar AR40),另一只眼植入锐边疏水丙烯酸人工晶状体(AMO Sensar OptiEdge AR40e)。第2组(26例患者)双眼均植入带有截短光学部的硅酮人工晶状体(Pharmacia CeeOn 911A)。在该组中,一只眼的前囊进行广泛抛光,另一只眼不进行抛光。术后1年使用标准化摄影技术拍摄数字化裂隙灯照片以观察纤维化PCO情况。由2名盲法检查者对PCO强度进行主观分级(0至4分)。
2名检查者之间主观PCO评分相关性良好(r = 0.88)。在第1组中,OptiEdge AR40e人工晶状体的平均PCO评分为0.26,AR40人工晶状体为0.90(P<0.01)。在第2组中,前囊抛光眼的平均PCO评分为0.24,未抛光眼为0.17(P = 0.31)。
术后1年,锐边的OptiEdge AR40e人工晶状体导致的周边纤维化PCO明显少于圆边的AR40人工晶状体。对于植入锐边硅酮911A人工晶状体的眼,前囊抛光在纤维化PCO方面未造成显著差异。