Baez K A, Orengo S, Gandham S, Spaeth G L
William & Anna Goldberg Glaucoma Service and Research Center, Wills Eye Hospital/Thomas Jefferson University, Philadelphia, Pa. 19107.
Ophthalmic Surg. 1992 Jun;23(6):426-8.
The NIDEK EAS-1000 is a camera and computer system designed to facilitate measurement and analysis of the anterior segment of the eye. To evaluate the interobserver reproducibility of the system, three examiners assessed the anterior chamber depth and peripheral angle measurement in 20 disease-free eyes. Intraobserver reproducibility was determined by each investigator's photographing and analyzing the same eye on three different occasions. Interobserver correlation for the anterior chamber depth (using the first three readings of the same eye made by each examiner) was 0.97, with an 95% CI of 0.94 to 0.98. Interobserver correlation for angle measurement (again using the first three readings of the same eye made by each examiner) was 0.89, with a 95% confidence interval (CI) of 0.81 to 0.95. Intraobserver correlations for anterior chamber depth were 0.98 for each of the three examiners (mean, 0.98), wiht a 95% CI of 0.91 to 0.99. Intraobserver correlations for anterior chamber depth were 0.98 for each of the three examiners (mean, 0.98), with a 95% CI of 0.91 to 0.99. The intraobserver CI for angle measurement was 0.97 (range, 0.95 to 0.99). The intraobserver and interobserver reproducibility of the NIDEK EAS-1000 system was excellent.
尼德克EAS - 1000是一种相机和计算机系统,旨在便于对眼前节进行测量和分析。为评估该系统在不同观察者之间的可重复性,三名检查者对20只无疾病的眼睛的前房深度和周边房角进行了测量。观察者自身的可重复性由每位研究者在三个不同时间对同一只眼睛进行拍照和分析来确定。前房深度的观察者间相关性(使用每位检查者对同一只眼睛的前三次读数)为0.97,95%置信区间为0.94至0.98。房角测量的观察者间相关性(同样使用每位检查者对同一只眼睛的前三次读数)为0.89,95%置信区间为0.81至0.95。三名检查者各自的前房深度观察者自身相关性均为0.98(平均值为0.98),95%置信区间为0.9至0.99。三名检查者各自的前房深度观察者自身相关性均为0.98(平均值为0.98),95%置信区间为0.9至0.99。房角测量的观察者自身置信区间为0.97(范围为0.95至0.99)。尼德克EAS - 1000系统在观察者自身和不同观察者之间的可重复性都非常好。