Gosheger Georg, Hardes Jendrik, Ahrens Helmut, Streitburger Arne, Buerger Horst, Erren Michael, Gunsel Andreas, Kemper Fritz H, Winkelmann Winfried, Von Eiff Christof
Department of Orthopedics, University of Muenster, Albert-Schweitzer-Str. 33, 48149 Muenster, Germany.
Biomaterials. 2004 Nov;25(24):5547-56. doi: 10.1016/j.biomaterials.2004.01.008.
Deep infection of megaprostheses remains a serious complication in orthopedic tumor surgery. Despite the use of systemic and local antibiotic prophylaxis the reported infection rate is between 5% and 35%. Silver-coated medical devices proved their effectiveness in reducing infections. The objective of this study was to examine in vivo the antimicrobial efficacy and possible side-effects of a silver-coated megaprosthesis. In a first study, 30 rabbits (15 titanium versus 15 silver-coated Mutars-endoprostheses) were infected with Staphylococcus aureus. In a second study, toxicological side effects were analyzed in 10 rabbits with a silver-coated megaprosthesis. The silver group showed significantly (p<0.05) lower infection rates (7% versus 47%) in comparison with the titanium group. Measurements of the C-reactive-protein, neutrophilic leukocytes, rectal temperature and body weight showed significant (p<0.05) lower signs of inflammation in the silver group. The analysis of the silver concentration in blood (median 1.883ppb) and in organs (0.798-86.002ppb) showed elevated silver concentrations without pathologic changes in laboratory parameters and without histological changes of organs. In conclusion, the new silver-coated Mutars-megaprosthesis resulted in reduced infection rates without toxicological side effects, suggesting that this prosthesis might be a promising device in tumor surgery exhibiting antimicrobial activity.
大假体深部感染仍然是骨肿瘤手术中的一种严重并发症。尽管使用了全身和局部抗生素预防措施,但报告的感染率仍在5%至35%之间。镀银医疗设备已证明其在减少感染方面的有效性。本研究的目的是在体内检测镀银大假体的抗菌效果和可能的副作用。在第一项研究中,30只兔子(15只植入钛制假体,15只植入镀银Mutars型内置假体)感染了金黄色葡萄球菌。在第二项研究中,对10只植入镀银大假体的兔子进行了毒理学副作用分析。与钛制假体组相比,镀银组的感染率显著降低(p<0.05)(7%对47%)。对C反应蛋白、中性粒细胞、直肠温度和体重的测量显示,镀银组的炎症迹象显著降低(p<0.05)。血液(中位数1.883ppb)和器官(0.798 - 86.002ppb)中银浓度的分析显示,银浓度升高,但实验室参数无病理变化,器官也无组织学变化。总之,新型镀银Mutars大假体降低了感染率,且无毒理学副作用,这表明该假体可能是肿瘤手术中一种具有抗菌活性的有前景的器械。