颈椎螺钉拔出强度、插入扭矩与骨密度关系的生物力学评估

Biomechanical evaluation of relationship of screw pullout strength, insertional torque, and bone mineral density in the cervical spine.

作者信息

Reitman Charles Alan, Nguyen Lyndon, Fogel Guy R

机构信息

Baylor College of Medicine, Houston, TX 77030, USA.

出版信息

J Spinal Disord Tech. 2004 Aug;17(4):306-11. doi: 10.1097/01.bsd.0000090575.08296.9d.

Abstract

BACKGROUND

Understanding of implant failure mechanisms is important in the successful utilization of anterior cervical plates. Many variables influence screw purchase, including the quality of the bone. The purpose of this study was to assess the relationship of screw pullout and screw insertional torque across a wide range of bone mineral densities (BMDs).

METHODS

A total of 54 cervical vertebrae in 12 cervical spines were evaluated for BMD using dual-energy x-ray absorptiometry scanning. Actual and perceived peak torques of 3.5-mm anterior cervical screws were determined at each level followed by screw pullout strength testing.

RESULTS

A high correlation was observed between screw pullout strength and BMD. However, there was a low correlation of peak insertional torque to pullout strength.

CONCLUSION

These findings suggest the quality of the bone is more instrumental in the success or failure of anterior cervical screws than is the insertional torque with which the screws are placed.

摘要

背景

了解植入物失败机制对于成功使用颈椎前路钢板至关重要。许多变量会影响螺钉的把持力,包括骨质质量。本研究的目的是评估在广泛的骨矿物质密度(BMD)范围内螺钉拔出力与螺钉插入扭矩之间的关系。

方法

使用双能X线吸收法扫描对12个颈椎的54个颈椎椎体进行骨密度评估。在每个节段测定3.5毫米颈椎前路螺钉的实际和感知峰值扭矩,随后进行螺钉拔出强度测试。

结果

观察到螺钉拔出强度与骨密度之间存在高度相关性。然而,峰值插入扭矩与拔出强度之间的相关性较低。

结论

这些发现表明,与螺钉置入时的插入扭矩相比,骨质质量在颈椎前路螺钉的成败中更具决定性作用。