Gardner R M, Crapo R O, Jackson B R, Jensen R L
LDS Hospital/University of Utah, Salt Lake City 84143.
Chest. 1992 Apr;101(4):948-52. doi: 10.1378/chest.101.4.948.
Peak flow meters provide physicians and patients with objective measures about changes in pulmonary obstruction. We evaluated eight models of peak flowmeters and measured their accuracy and reproducibility with methods recently recommended by the National Asthma Education Program (NAEP). Waveforms from the American Thoracic Society's spirometer testing set were used to drive a computer-controlled syringe. Testing was done at Salt Lake City at an altitude 1,400 m. It appears that the original Wright peak flowmeter has been used as the "de facto" standard. We found that the original Wright peak flowmeter overestimated flows in its midrange; and, as a consequence, most of the other peak flowmeters also overestimated peak flows. The overestimation of peak flows may have been understated because of the 1,400-m altitude testing site. To the credit of the instrument manufacturers, we were pleasantly surprised with the quality, accuracy, and reproducibility of presently available peak flowmeters; however, as a result of our testing, we suspect that with little effort, manufacturers of peak flowmeters could improve the accuracy of their devices. Standardized testing methods and equipment should make the task of peak flowmeter design, manufacture, and testing even easier. We trust that manufacturers of peak flowmeters will respond appropriately and improve their instruments.
峰值流量计可为医生和患者提供有关肺阻塞变化的客观测量数据。我们评估了八种型号的峰值流量计,并采用美国国家哮喘教育计划(NAEP)最近推荐的方法测量了它们的准确性和可重复性。美国胸科学会肺活量计测试装置的波形被用于驱动一个计算机控制的注射器。测试在盐湖城海拔1400米处进行。似乎原来的赖特峰值流量计一直被用作“事实上的”标准。我们发现原来的赖特峰值流量计在其流量范围的中间部分高估了流量;因此,大多数其他峰值流量计也高估了峰值流量。由于测试地点在海拔1400米处,峰值流量的高估可能被低估了。值得称赞的是仪器制造商,目前可用的峰值流量计的质量、准确性和可重复性让我们感到惊喜;然而,根据我们的测试,我们怀疑峰值流量计制造商只需付出一点努力就能提高其设备的准确性。标准化的测试方法和设备应该会使峰值流量计的设计、制造和测试任务变得更加容易。我们相信峰值流量计制造商将会做出适当回应并改进他们的仪器。