Thermal changes and cure depths associated with a high intensity light activation unit.

作者信息

Stewardson D A, Shortall A C C, Harrington E, Lumley P J

机构信息

The School of Dentistry, The University of Birmingham, St Chad's Queensway, Birmingham B4 6NN, UK.

出版信息

J Dent. 2004 Nov;32(8):643-51. doi: 10.1016/j.jdent.2004.06.007.

Abstract

OBJECTIVES

The objectives of this study were to evaluate the thermal emission and curing characteristics of a high intensity halogen light (Astralis 10-Ivoclar Vivadent, Schaan. Leichtenstein) alone and on curing a conventional and a fast-curing micro-hybrid composite.

METHODS

A bead thermistor was placed in the base of a standard model cavity. The cavity was irradiated using the light unit whilst empty, and when filled with either composite. Temperature rises were recorded using the light in four different output modes. Further samples were prepared to assess depth of cure via a digital penetrometer, light transmission using a computer-based radiometer, and microhardness with a Wallace hardness tester.

RESULTS

Mean peak temperature rises recorded during polymerisation of the composites ranged from 6.9 degrees C for the product InTen-S cured with the Adhesive programme (Adh) to 11.0 degrees C for the product Tetric Ceram HB cured with the High Power (HI P) programme. A significantly greater depth of cure was obtained for InTen-S in line with the greater light transmission obtained for this material.

DISCUSSION

The very high thermal emission characteristics reported in a recently published investigation for this light unit were not confirmed.

CONCLUSIONS

As the extent of thermal trauma that can be tolerated by the dental pulp is unknown consideration should be given to the choice of light activation unit and curing programme when polymerising light activated resin based restorations in deep cavities close to the pulp.