Suppr超能文献

The porous-coated anatomic total hip prosthesis: 11- to 13-year results.

作者信息

Moskal Joseph T, Jordan Louis, Brown Thomas E

机构信息

Roanoke Orthopaedic Center, Virginia 24018, USA.

出版信息

J Arthroplasty. 2004 Oct;19(7):837-44. doi: 10.1016/j.arth.2004.01.004.

Abstract

Of 137 consecutive primary porous-coated anatomic (PCA) arthroplasties performed, complete data was available on 107 hips in 93 patients at an average of 12.4 years (range, 11-13 years). The average Harris Hip Score improved from 44 to 85 at final follow-up, with 78% good or excellent result. Thirteen percent (14 of 107) of acetabular components and 4 porous ingrowth stems were deemed failures. Osteolytic lesions were seen in 7.5% (8 of 107) of acetabular components. Survivorship of the PCA prosthesis as a whole was 83% at 12.4 years. This first-generation femoral stem performed well, with only 4 failures, all of which occurred in patients with avascular necrosis in which unrecognized pathologic changes in the proximal femur may have inhibited bone ingrowth. The majority of the failures can be attributed to aseptic loosening of the acetabular component.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验