Caliendo Angela M, Ingersoll Jessica, Green Alicia M, Nolte Frederick S, Easley Kirk A
Department of Pathology and Laboratory Medicine, Emory University, Atlanta, GA, USA.
J Clin Microbiol. 2004 Nov;42(11):5392-3. doi: 10.1128/JCM.42.11.5392-5393.2004.
This study compared the sensitivity and viral load values of the AMPLICOR HIV-1 MONITOR microwell version 1.0, microwell version 1.5, and COBAS version 1.5 tests. Based on the percentage of positive replicates, the microwell version 1.5 and COBAS version 1.5 tests are more sensitive than the microwell version 1.0 test. Viral load values obtained with the COBAS version 1.5 test are lower than those obtained with either the microwell version 1.0 or microwell version 1.5 test.
本研究比较了AMPLICOR HIV-1 MONITOR微孔板1.0版、微孔板1.5版和COBAS 1.5版检测的灵敏度和病毒载量值。基于阳性重复检测的百分比,微孔板1.5版和COBAS 1.5版检测比微孔板1.0版检测更灵敏。COBAS 1.5版检测获得的病毒载量值低于微孔板1.0版或微孔板1.5版检测获得的值。