An automated spectrophotometric method for measuring canine ceruloplasmin in serum.

作者信息

Cerón Jose Joaquin, Martínez-Subiela Silvia

机构信息

Department of Animal Medicine and Surgery, Veterinary School, University of Murcia, 30100 Espinardo, Murcia, Spain.

出版信息

Vet Res. 2004 Nov-Dec;35(6):671-9. doi: 10.1051/vetres:2004046.

Abstract

An automated method for the determination of ceruloplasmin activity was developed and validated in canine serum. The method is based on the in vitro oxidase activity that this protein shows with substances such as p-phenylenediamine. In order to determine optimum assay conditions, the effects of the substrate concentration, buffer pH, reaction time and EDTA on the reaction were evaluated. The precision of the assay was good with within-run and between-run coefficients of variation lower than 10%. The method measured the ceruloplasmin values in a proportional and linear manner (r = 0.99) with a limit of detection of 0.0007 +/- 0.0001 Delta Abs/min. A temperature of -20 degrees C kept the reagent stable for 30 days. The method is cheap and easy to adapt to any automated biochemical analyser, considerably decreasing the processing time required with the manual method. Additionally it allows to differentiate dogs with pyometra and trauma from clinically healthy dogs.

摘要

开发了一种用于测定犬血清中铜蓝蛋白活性的自动化方法并进行了验证。该方法基于这种蛋白质与对苯二胺等物质显示的体外氧化酶活性。为了确定最佳测定条件,评估了底物浓度、缓冲液pH值、反应时间和EDTA对反应的影响。该测定的精密度良好,批内和批间变异系数低于10%。该方法以比例和线性方式(r = 0.99)测量铜蓝蛋白值,检测限为0.0007 +/- 0.0001吸光度变化/分钟。-20℃的温度可使试剂稳定30天。该方法成本低廉,易于适配任何自动化生化分析仪,大大减少了手工方法所需的处理时间。此外,它还能区分患有子宫蓄脓和创伤的犬与临床健康犬。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验