Suppr超能文献

使用多回波EPI进行图像失真的单次补偿和BOLD对比度优化以实现实时功能磁共振成像

Single-shot compensation of image distortions and BOLD contrast optimization using multi-echo EPI for real-time fMRI.

作者信息

Weiskopf Nikolaus, Klose Uwe, Birbaumer Niels, Mathiak Klaus

机构信息

Institute of Medical Psychology and Behavioral Neurobiology, University of Tübingen, Tübingen, Germany.

出版信息

Neuroimage. 2005 Feb 15;24(4):1068-79. doi: 10.1016/j.neuroimage.2004.10.012. Epub 2004 Dec 8.

Abstract

Functional magnetic resonance imaging (fMRI) is most commonly based on echo-planar imaging (EPI). With higher field strengths, gradient performance, and computational power, real-time fMRI has become feasible; that is, brain activation can be monitored during the ongoing scan. However, EPI suffers from geometric distortions due to inhomogeneities of the magnetic field, especially close to air-tissue interfaces. Thus, functional activations might be mislocalized and assigned to the wrong anatomical structures. Several techniques have been reported which reduce geometric distortions, for example, mapping of the static magnetic field B(0) or the point spread function for all voxels. Yet these techniques require additional reference scans and in some cases extensive computational time. Moreover, only static field inhomogeneities can be corrected, because the correction is based on a static reference scan. We present an approach which allows for simultaneous acquisition and distortion correction of a functional image without a reference scan. The technique is based on a modified multi-echo EPI data acquisition scheme using a phase-encoding (PE) gradient with alternating polarity. The images exhibit opposite distortions due to the inverted PE gradient. After adjusting the contrast of the images acquired at different echo times, this information is used for the distortion correction. We present the theory, implementation, and applications of this single-shot distortion correction. Significant reduction in geometric distortion is shown both for phantom images and human fMRI data. Moreover, sensitivity to the blood oxygen level-dependent (BOLD) effect is increased by weighted summation of the undistorted images.

摘要

功能磁共振成像(fMRI)最常基于回波平面成像(EPI)。随着更高的场强、梯度性能和计算能力,实时fMRI已变得可行;也就是说,可以在正在进行的扫描过程中监测大脑激活情况。然而,由于磁场的不均匀性,尤其是在靠近空气 - 组织界面处,EPI会受到几何失真的影响。因此,功能激活可能会被错误定位并归因于错误的解剖结构。已经报道了几种减少几何失真的技术,例如,对所有体素的静磁场B(0)或点扩散函数进行映射。然而,这些技术需要额外的参考扫描,并且在某些情况下需要大量的计算时间。此外,由于校正是基于静态参考扫描,所以只能校正静态场不均匀性。我们提出了一种方法,该方法允许在无需参考扫描的情况下同时采集和校正功能图像的失真。该技术基于一种改进的多回波EPI数据采集方案,使用具有交替极性的相位编码(PE)梯度。由于PE梯度反转,图像呈现相反的失真。在调整了在不同回波时间采集的图像的对比度后,此信息用于失真校正。我们展示了这种单次失真校正的理论、实现和应用。对于体模图像和人类fMRI数据,几何失真都有显著降低。此外,通过对未失真图像进行加权求和,提高了对血氧水平依赖(BOLD)效应的敏感性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验