Bourel P, Gibon D, Coste E, Daanen V, Rousseau J
Institute of Medical Technology, University Hospital of Lille, France.
Med Phys. 1999 Dec;26(12):2693-700. doi: 10.1118/1.598809.
The authors have developed a protocol and software for the quality assessment of MRI equipment with a commercial test object. Automatic image analysis consists of detecting surfaces and objects, defining regions of interest, acquiring reference point coordinates and establishing gray level profiles. Signal-to-noise ratio, image uniformity, geometrical distortion, slice thickness, slice profile, and spatial resolution are checked. The results are periodically analyzed to evaluate possible drifts with time. The measurements are performed weekly on three MRI scanners made by the Siemens Company (VISION 1.5T, EXPERT 1.0T, and OPEN 0.2T). The results obtained for the three scanners over approximately 3.5 years are presented, analyzed, and compared.
作者们开发了一种使用商业测试对象对MRI设备进行质量评估的方案和软件。自动图像分析包括检测表面和物体、定义感兴趣区域、获取参考点坐标以及建立灰度剖面图。检查信噪比、图像均匀性、几何畸变、切片厚度、切片轮廓和空间分辨率。定期分析结果以评估随时间可能出现的漂移。每周在三台西门子公司生产的MRI扫描仪(VISION 1.5T、EXPERT 1.0T和OPEN 0.2T)上进行测量。展示、分析并比较了这三台扫描仪在大约3.5年时间里获得的结果。