Comparison of the stability of 1-piece and 3-piece acrylic intraocular lenses in the lens capsule.

作者信息

Hayashi Ken, Hayashi Hideyuki

机构信息

Hayashi Eye Hospital, Fukuoka, Japan.

出版信息

J Cataract Refract Surg. 2005 Feb;31(2):337-42. doi: 10.1016/j.jcrs.2004.06.042.

Abstract

PURPOSE

To compare the degrees of intraocular lens (IOL) decentration, tilt, and longitudinal movement; the refractive change; and anterior capsule contraction after cataract surgery between eyes with a 1-piece acrylic IOL with soft acrylic loops and eyes with a 3-piece acrylic IOL with rigid poly(methyl methacrylate) loops.

SETTING

Hayashi Eye Hospital, Fukuoka, Japan.

METHODS

Fifty-six patients had implantation of a 1-piece acrylic IOL in 1 eye and a 3-piece acrylic IOL in the fellow eye. The degree of IOL decentration and tilt, the anterior chamber depth (ACD), and the area of anterior capsule opening were measured using Scheimpflug videophotography 3 days and 1, 3, and 6 months postoperatively. The postoperative refractive status was also examined.

RESULTS

The mean degrees of decentration and tilt in the 1-piece IOL group were similar to those in the 3-piece IOL group throughout the follow-up. The ACD did not change after surgery in the 1-piece group but showed significant shallowing in the 3-piece group (P < .0001). The spherical equivalent (SE) did not change in the 1-piece group, while the SE had a significant myopic shift of approximately 0.4 diopter in the 3-piece group. The percentage of anterior capsule contraction was similar between the groups.

CONCLUSIONS

The degree of IOL decentration and tilt and percentage of anterior capsule contraction in eyes with a 1-piece acrylic IOL with soft acrylic loops were similar to those in eyes with a 3-piece acrylic IOL. The longitudinal movement of the 1-piece IOL was less than the movement of the 3-piece IOL, resulting in less postoperative myopic shift.

摘要

目的

比较采用带有柔软丙烯酸襻的单片式丙烯酸人工晶状体(IOL)的眼与采用带有硬质聚甲基丙烯酸甲酯襻的三片式丙烯酸IOL的眼在白内障手术后人工晶状体的偏心、倾斜和纵向移动程度;屈光变化;以及前囊收缩情况。

设置

日本福冈市林眼科医院。

方法

56例患者一只眼植入单片式丙烯酸IOL,另一只眼植入三片式丙烯酸IOL。在术后3天以及1、3和6个月时,使用Scheimpflug摄像测量人工晶状体的偏心和倾斜程度、前房深度(ACD)以及前囊开口面积。还检查了术后屈光状态。

结果

在整个随访过程中,单片式IOL组的平均偏心和倾斜程度与三片式IOL组相似。单片式IOL组术后ACD未改变,而三片式IOL组ACD显著变浅(P < .0001)。单片式IOL组的等效球镜度(SE)未改变,而三片式IOL组的SE有大约0.4屈光度的显著近视性偏移。两组间前囊收缩百分比相似。

结论

采用带有柔软丙烯酸襻的单片式丙烯酸IOL的眼与采用三片式丙烯酸IOL的眼的人工晶状体偏心和倾斜程度以及前囊收缩百分比相似。单片式IOL的纵向移动小于三片式IOL,导致术后近视性偏移较小。