Båth Magnus, Håkansson Markus, Hansson Jonny, Månsson Lars Gunnar
Department of Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, SE-413 45 Göteborg, Sweden.
Radiat Prot Dosimetry. 2005;114(1-3):230-5. doi: 10.1093/rpd/nch567.
Using a completely digital environment for the entire imaging process leads to new possibilities for optimisation of radiography since many restrictions of screen/film systems, such as the small dynamic range and the lack of possibilities for image processing, do not apply any longer. However, at the same time these new possibilities lead to a more complicated optimisation process, since more freedom is given to alter parameters. This paper focuses on describing an optimisation strategy that concentrates on taking advantage of the conceptual differences between digital systems and screen/film systems. The strategy can be summarised as: (a) always include the anatomical background during the optimisation, (b) perform all comparisons at a constant effective dose and (c) separate the image display stage from the image collection stage. A three-step process is proposed where the optimal setting of the technique parameters is determined at first, followed by an optimisation of the image processing. In the final step the optimal dose level-given the optimal settings of the image collection and image display stages-is determined.
在整个成像过程中使用完全数字化的环境为X射线摄影的优化带来了新的可能性,因为屏/片系统的许多限制,如动态范围小和缺乏图像处理的可能性,已不再适用。然而,与此同时,这些新的可能性导致了一个更加复杂的优化过程,因为在改变参数方面有了更多的自由度。本文着重描述一种优化策略,该策略专注于利用数字系统和屏/片系统之间的概念差异。该策略可概括为:(a) 在优化过程中始终纳入解剖学背景,(b) 在恒定有效剂量下进行所有比较,以及(c) 将图像显示阶段与图像采集阶段分开。提出了一个三步过程,首先确定技术参数的最佳设置,然后进行图像处理的优化。在最后一步中,在给定图像采集和图像显示阶段的最佳设置的情况下,确定最佳剂量水平。