使用BOD POD进行脂肪百分比估计时的设备间变异性。

Interdevice variability in percent fat estimates using the BOD POD.

作者信息

Ball S D

机构信息

Department of Nutritional Sciences, University of Missouri-Columbia, Columbia, MO, USA.

出版信息

Eur J Clin Nutr. 2005 Sep;59(9):996-1001. doi: 10.1038/sj.ejcn.1602202.

Abstract

OBJECTIVE

To evaluate interdevice reliability in body density (Db) and percent body fat (%BF) using air-displacement plethysmography, the BOD POD (BP) body composition system.

DESIGN AND SETTING

Duplicate body composition tests were performed in immediate succession on 50 adults (26 M, 24 F; 21-53 y) using two BP units located in the same body composition laboratory.

RESULTS

Mean Db and %BF between BP1 and BP2 did not differ significantly for men (DeltaDb = 0.0003+/-0.0008 g/ml, P = 0.632; Delta%BF = 0.1+/-1.3, P = 0.665), while for women, there were small but significant differences in Db and %BF between BP1 and BP2 (DeltaDb = 0.0018+/-0.0003 g/ml, P = 0.001; Delta%BF = 0.8+/-1.1, P = 0.001). The regression between %BF by BP1 and BP2 did not deviate significantly from the line of identity for both men and women (R2 = 0.95, standard error of estimate (s.e.e.) = 1.23 %BF for men; R2 = 0.97, s.e.e = 1.13 %BF for women). Individual variations in %BF estimates between the two BP units were within acceptable ranges (95% limits of agreement = -2.5-2.7 %BF for men; -1.4-3.0 %BF for women), and there was no trend in individual differences as %BF varied (r = -0.19, P = 0.359 for men; r = 0.09, P = 0.677 for women). Other subject characteristics, including age, body mass, height, and body mass index, did not significantly contribute to the differences in %BF estimates by the two BP units.

CONCLUSIONS

No clinically significant differences in Db and %BF estimates exist between the BP units, and the interdevice variability of the BP has minimal impact on %BF estimates. Further, test-to-test reliability between BP units appears to be as good as within one unit.

摘要

目的

使用空气置换体积描记法(BOD POD,BP)人体成分系统评估不同设备在身体密度(Db)和体脂百分比(%BF)方面的可靠性。

设计与环境

在同一人体成分实验室中,使用两台BP设备对50名成年人(26名男性,24名女性;年龄21 - 53岁)连续立即进行重复的人体成分测试。

结果

对于男性,BP1和BP2之间的平均Db和%BF无显著差异(ΔDb = 0.0003±0.0008 g/ml,P = 0.632;Δ%BF = 0.1±1.3,P = 0.665),而对于女性,BP1和BP2之间的Db和%BF存在小但显著的差异(ΔDb = 0.0018±0.0003 g/ml,P = 0.001;Δ%BF = 0.8±1.1,P = 0.001)。BP1和BP2测得的%BF之间的回归对于男性和女性均未显著偏离恒等线(男性R2 = 0.95,估计标准误(s.e.e.)= 1.23%BF;女性R2 = 0.97,s.e.e = 1.13%BF)。两个BP设备之间%BF估计值的个体差异在可接受范围内(男性95%一致性界限 = -2.5 - 2.7%BF;女性 -1.4 - 3.0%BF),并且随着%BF变化,个体差异无趋势(男性r = -0.19,P = 0.359;女性r = 0.09,P = 0.677)。其他受试者特征,包括年龄、体重、身高和体重指数,对两个BP设备%BF估计值的差异无显著影响。

结论

BP设备之间在Db和%BF估计值方面不存在临床上显著的差异,并且BP设备间的变异性对%BF估计值的影响最小。此外,BP设备之间的测试 - 测试可靠性似乎与同一设备内的可靠性一样好。