在1.5-T磁共振扫描仪内使用自动系统进行磁共振引导下大芯针乳腺活检:体外实验及4例患者的初步临床经验

Magnetic resonance-guided large-core breast biopsy inside a 1.5-T magnetic resonance scanner using an automatic system: in vitro experiments and preliminary clinical experience in four patients.

作者信息

Pfleiderer Stefan O R, Marx Christiane, Vagner Jörg, Franke Ralf-Peter, Reichenbach Jürgen R, Kaiser Werner A

机构信息

Institute of Diagnostic and Interventional Radiology, Friedrich-Schiller-University Jena, Jena, Germany.

出版信息

Invest Radiol. 2005 Jul;40(7):458-63. doi: 10.1097/01.rli.0000167423.27180.54.

Abstract

RATIONALE AND OBJECTIVE

The aim of this study was to investigate the feasibility and the precision of magnetic resonance (MR)-guided large-core breast biopsies (LCBB) by using the second prototype of an automatic system (ROBITOM II), which is used to localize lesions while operating at the isocenter of a 1.5-T whole-body scanner.

METHODS AND MATERIALS

In comparison to the first prototype, ROBITOM II is equipped with a dedicated double breast coil and a high-speed trocar setting unit. In vitro experiments (n = 25) with grapefruit phantoms, which contained multiple vitamin E capsules (12 x 7 mm in size) as artificial lesions, were performed. Four patients with MR-detectable breast lesions underwent biopsy. A trocar was positioned in front of the lesion and inserted into the breast. Specimens were harvested with a coaxial technique by using a 14-G core needle biopsy gun.

RESULTS

In all 25 in vitro experiments, capsule material was detected in the specimen cylinder. In 4 patients, the coaxial needle was detected exactly at the expected position. Between 8 and 16 tissue cylinders were harvested. Histologic evaluation resulted in 1 invasive ductal carcinoma and 1 papilloma, which were confirmed after open surgery. One patient who had a proven breast cancer was biopsied for exclusion of multifocal disease. She showed fibrocystic changes, whereas open surgery revealed 3 small areas of ductal carcinoma in situ (DCIS). Another patient showed fibroadenoma after biopsy. This patient is in the follow-up period, which has lasted between 3 and 4 months up until now.

CONCLUSIONS

In this pilot patient study, the feasibility of manipulator-assisted large-core breast biopsy inside a 1.5-T whole-body scanner was demonstrated by using ROBITOM II. The precision of the device was confirmed with in vitro experiments. Although these findings are preliminary and the follow-up period is rather short, they nevertheless represent a successful proof-of-principle of LCBB with ROBITOM II.

摘要

原理与目的

本研究旨在通过使用自动系统的第二代原型(ROBITOM II)来研究磁共振(MR)引导下大芯针乳腺活检(LCBB)的可行性和精确性,该系统用于在1.5-T全身扫描仪的等中心操作时定位病变。

方法与材料

与第一代原型相比,ROBITOM II配备了专用的双乳腺线圈和高速套管针设置单元。对含有多个维生素E胶囊(尺寸为12×7毫米)作为人工病变的葡萄柚模型进行了25次体外实验。4例有MR可检测乳腺病变的患者接受了活检。将套管针置于病变前方并插入乳腺。通过使用14-G芯针活检枪采用同轴技术获取标本。

结果

在所有25次体外实验中,标本筒中均检测到胶囊材料。在4例患者中,同轴针被准确检测到预期位置。获取了8至16个组织筒。组织学评估结果为1例浸润性导管癌和1例乳头状瘤,开放手术后得到证实。1例确诊为乳腺癌的患者接受活检以排除多灶性疾病。她显示为纤维囊性改变,而开放手术发现3个小的原位导管癌(DCIS)区域。另1例患者活检后显示为纤维腺瘤。该患者目前处于随访期,截至目前已持续3至4个月。

结论

在这项初步的患者研究中,通过使用ROBITOM II证明了在1.5-T全身扫描仪内进行机械手辅助大芯针乳腺活检的可行性。该设备的精确性通过体外实验得到证实。尽管这些发现是初步的且随访期较短,但它们仍然代表了使用ROBITOM II进行LCBB的成功原理验证。