Griffin P M, Zamel N
J Appl Physiol Respir Environ Exerc Physiol. 1979 Nov;47(5):1127-30. doi: 10.1152/jappl.1979.47.5.1127.
A modification of the conventional volume-displacement whole-body plethysmograph was done using a low-resistance flow element and integrating the plethysmographic pressure signal. A flow element with a resistance of 0.036 cmH2O/(l/s), linear to 75 l/s was fitted to a 440-liter plethysmograph, resulting in an instrument with a volume-amplitude-frequency response flat to 15 Hz and in phase with mouth flow and esophageal catheter-balloon to at least 15 Hz without pressure compensation.