Comparison of different body size parameters for individual dose adaptation in body CT of adults.

作者信息

Menke Jan

机构信息

Department of Diagnostic Radiology, University Hospital, Rosenwinkel 5, 37081 Goettingen, Germany.

出版信息

Radiology. 2005 Aug;236(2):565-71. doi: 10.1148/radiol.2362041327.

Abstract

PURPOSE

To investigate prospectively which of several body size parameters are suitable for individual dose adaptation in body computed tomography (CT) of adults.

MATERIALS AND METHODS

Three body regions (thorax, abdomen, pelvis) were scanned exclusively for clinical reasons, with institutional ethical approval and informed consent. For each of the three regions, 50 men and 50 women (aged 18-87 years) were studied (300 scans total). Individual x-ray properties for each scan were summarized with a water-equivalent diameter (Dw). Different body size parameters, based on weight, height, and shape, were correlated with Dw by using regression analysis. This resulted in Dw estimation errors of different magnitudes, indicated with 95% prediction intervals. The errors from weight were compared with those from each of the other body parameters by using comparison of variance in paired samples (P < .05). In addition, a topogram-based estimate for Dw was studied, which simulated an automated body size measurement.

RESULTS

For the thorax, abdomen, and pelvis, mean Dw was 28.0, 29.1, and 29.3 cm, and estimation of Dw from weight enabled 95% prediction intervals of +/-2.5, +/-2.4, and +/-2.6 cm, respectively. Combinations of height and weight were only slightly more or even less exact than were measurements from only weight. Diameter-related parameters such as body circumference were similar to or better than weight. However, the topogram-based estimate was significantly more exact.

CONCLUSION

Body weight and circumference enable suitable estimates for individual dose adaptation in body CT of adults if automated dose adaptation is not available.

摘要

目的

前瞻性地研究在成人身体计算机断层扫描(CT)中,几种身体尺寸参数中哪些适用于个体剂量调整。

材料与方法

仅出于临床原因,在获得机构伦理批准并取得知情同意的情况下,对三个身体区域(胸部、腹部、骨盆)进行扫描。对这三个区域中的每一个区域,研究了50名男性和50名女性(年龄18 - 87岁)(共300次扫描)。每次扫描的个体X射线特性用水等效直径(Dw)进行总结。基于体重、身高和体型的不同身体尺寸参数通过回归分析与Dw相关联。这导致了不同量级的Dw估计误差,并以95%预测区间表示。通过配对样本方差比较(P <.05),将体重产生的误差与其他每个身体参数产生的误差进行比较。此外,还研究了基于定位像的Dw估计方法,该方法模拟了自动身体尺寸测量。

结果

对于胸部、腹部和骨盆,平均Dw分别为28.0、29.1和29.3厘米,由体重估计Dw时,其95%预测区间分别为±2.5、±2.4和±2.6厘米。身高和体重的组合仅比仅用体重测量稍微精确一点或甚至更不精确。与直径相关的参数,如身体周长,与体重相似或比体重更好。然而,基于定位像的估计明显更精确。

结论

如果没有自动剂量调整,体重和身体周长可为成人身体CT中的个体剂量调整提供合适的估计。