In a quasi-simultaneous assessment, imprecise cholesterol monitoring and screening tests were improved.

作者信息

Costanza Michael C, Wolff Hans, James Richard W, Morabia Alfredo

机构信息

Division of Clinical Epidemiology, Geneva University Hospitals, Switzerland.

出版信息

J Clin Epidemiol. 2005 Aug;58(8):841-8. doi: 10.1016/j.jclinepi.2005.01.009.

Abstract

OBJECTIVE

To calibrate Reflotron-measured capillary total cholesterol (cTC) relative to a laboratory-measured venous total cholesterol (vTC) criterion standard for monitoring and screening hypercholesterolemia.

STUDY DESIGN AND SETTING

Quasi-simultaneous assessment in 1999-2002 of Reflotron cTC and laboratory vTC in a random sample of 4,269 adult residents of Geneva, Switzerland (calibration development subsample n=3,067; validation subsample n=1,172), by means of bias, precision, correlation, sensitivity, and false positive percentage (calculated as 100-specificity) analyses of Reflotron cTC vs. laboratory vTC measures for predicting hypercholesterolemia.

RESULTS

Total bias was small (-0.26 mmol/L), but there was substantial negative drift in Reflotron cTC (annual biases +0.08, -0.17, -0.27, and -0.60 mmol/L in 1999-2002). Overall, 57% of Reflotron cTC measurements for 894 hypercholesterolemic patients underestimated laboratory vTC (2%, 57%, 71%, and 98% in 1999-2002). Prior to calibration, sensitivity was 73% for the development and 35% for the validation subsample, with false positive at 4% (development) and 0.1% (validation). After calibration, sensitivity was 78% for the development and 37% for the validation subsample, with false positive at 5% (development) and 0.2% (validation). Using 95% upper prediction limits (UPL) for individual vTC values increased sensitivity to 99% and 83% and false positive percentage to 30% and 7% for the development and validation subsamples, respectively.

CONCLUSION

Crude results of Reflotron-measured cTC have poor sensitivity. Instead, 95% UPL can be used for monitoring and screening. Simply adding 0.8 mmol/L to a patient's observed Reflotron cTC value provides a very good approximation to the 95% UPL.