Reflotron作为血液胆固醇筛查工具的有效性。

The validity of the Reflotron as a screening tool for blood cholesterol.

作者信息

Al-Humaidi M A, Abolfotouh M A, Sulaiman S A, Al-Kadoumi O F, Khattab M S, Al-Salmi H H, Al-Amari O, Saydain G, Al-Knawy B

机构信息

Department of Internal Medicine, King Saud University, College of Medicine, Abha, Kingdom of Saudi Arabia.

出版信息

J Egypt Public Health Assoc. 1997;72(1-2):167-87.

Abstract

Screening for blood cholesterol levels has become an essential need in many countries. The aim of this study was to evaluate the performance of Reflotron analyzer in screening for blood cholesterol in terms of its precision and accuracy. We have compared the results obtained by the Reflotron with those obtained by a standard hospital laboratory procedure for the same samples. An optimal precision that met the 1992 LSP standards for precision was observed (CV = 2.27%). On the other hand, a suboptimal accuracy of Reflotron was detected, where more than 80% of the Reflotron values differed from the reference laboratory values by more than 5%. As a consequence, 21% of individuals were misclassified as normal due to the very low sensitivity in spite of the optimum specificity observed for the Reflotron. Using the receiver operating characteristic curve analysis, the optimum cut-off points giving the best validity values for the test were 170 mg/dl and 190 mg/dl (4.4 mmol/L and 4.9 mmol/L) for the border-line and true hypercholesterolemia, respectively. At these levels, sensitivity was 0.89 and 0.86, specificity 0.93 and 0.93, and positive predictive value was 0.91 and 0.82, respectively. Considering some limitations of the study, it is concluded that the Reflotron could be of more potential validity for the use in cholesterol measurement surveys when the recommended cut-off points are used.

摘要

在许多国家,筛查血液胆固醇水平已成为一项基本需求。本研究的目的是评估Reflotron分析仪在筛查血液胆固醇方面的精密度和准确性。我们将Reflotron分析仪对相同样本的检测结果与标准医院实验室检测程序的结果进行了比较。观察到其精密度达到了1992年LSP精密度标准(变异系数=2.27%)。另一方面,检测发现Reflotron分析仪的准确性欠佳,超过80%的Reflotron检测值与参考实验室值相差超过5%。结果,尽管Reflotron分析仪的特异性最佳,但由于其灵敏度极低,21%的个体被误判为正常。通过绘制受试者工作特征曲线分析,对于临界高胆固醇血症和真正的高胆固醇血症,能给出最佳有效性值的最佳临界点分别为170mg/dl和190mg/dl(4.4mmol/L和4.9mmol/L)。在这些水平下,灵敏度分别为0.89和0.86,特异性分别为0.93和0.93,阳性预测值分别为0.91和0.82。考虑到本研究的一些局限性,得出的结论是,当使用推荐的临界点时,Reflotron分析仪在胆固醇测量调查中可能具有更大的潜在有效性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验