The standard pendulum appliance was modified by integrating a distal screw into its base and by special preactivation of the pendulum springs. The suitability of this Pendulum K for the translatory distalization of maxillary molars was investigated in an in vitro analysis and in an in vivo study in children and adolescents. The in vitro measurement of the resulting force systems revealed that all forces and moments remained virtually constant over a three-mm simulated distalization increment. The transverse force, Fx, increased from two to 11 cN and the weakly intrusively acting force, Fy, from six to eight cN, but these increases were not statistically significant. The distalization force, Fz, initially 201 cN, was still 199 cN after a three-mm distalization increment. The mesially acting moment, My, rose from 1654 to 1834 cN mm, whereas the palatally acting moment, Mz, declined slightly from 229 to 164 cN mm. The slight, consistent distoinclinatory moment, Mx, initially 306 cN mm, was 310 cN mm after three mm. In parallel, the in vivo study with its collective of 66 patients confirmed that the Pendulum K allows a virtually translatory molar distalization with slight tippings of 4.75 degrees to the palatal plane and 4.25 degrees to the anterior basal plane. Palatal movements of the first molars were avoided. The proportion of molar distalization in the total movement was 73.53%.
标准摆式矫治器通过在其基座中集成一个远中螺钉以及对摆簧进行特殊预激活来进行改良。在一项体外分析以及一项针对儿童和青少年的体内研究中,对这种改良后的摆式矫治器K用于上颌磨牙远中平移的适用性进行了研究。对所产生的力系统进行的体外测量显示,在模拟远中移动3毫米的增量过程中,所有的力和力矩实际上都保持恒定。横向力Fx从2厘牛顿增加到11厘牛顿,微弱的侵入性作用力Fy从6厘牛顿增加到8厘牛顿,但这些增加在统计学上并不显著。远中移动力Fz最初为201厘牛顿,在远中移动3毫米增量后仍为199厘牛顿。近中作用力矩My从1654厘牛顿·毫米增加到1834厘牛顿·毫米,而腭向作用力矩Mz从229厘牛顿·毫米略有下降至164厘牛顿·毫米。轻微的、持续的远中倾斜力矩Mx最初为306厘牛顿·毫米,在远中移动3毫米后为310厘牛顿·毫米。与此同时,对66名患者进行的体内研究证实,摆式矫治器K能够实现几乎是平移的磨牙远中移动,磨牙向腭平面轻微倾斜4.75度,向前基平面轻微倾斜4.25度。避免了第一磨牙的腭向移动。磨牙远中移动在总移动中的比例为73.53%。