SADJ. 2005 Jul;60(6):238-40.

Abstract

With water fluoridation imminent in South Africa, the accurate determination of the fluoride content of water is important. The aim of this study was to compare the fluoride content of water reported by 9 laboratories and the laboratory at the South African Bureau of Standards (SABS). The SABS and 9 South African laboratories were asked to determine the fluoride content of five water samples. The fluoride content of the samples was in a range that could be expected in South African waters. The laboratories were requested to disclose their methods for fluoride determination. The results reported by the laboratories were compared to the results reported by the SABS laboratory. Fluoride concentrations of 0.13, 0.22 and 0.58 mg/litre were reproduced to within 0.05 mg/litre by two, six and three laboratories respectively. At the 1.1 and 1.5 mg/litre concentration no laboratory could achieve this accuracy. Four different methods for the determination of fluoride were used. At present laboratories determining fluoride concentrations are not accurate enough to ensure that the process of water fluoridation will be safe. Laboratories will have to check their procedures to ensure better results before water fluoridation can commence.

摘要

随着南非即将实施水氟化,准确测定水中氟化物含量至关重要。本研究的目的是比较9个实验室以及南非标准局(SABS)实验室报告的水氟化物含量。要求SABS和9个南非实验室测定5个水样的氟化物含量。样品中的氟化物含量处于南非水域预期的范围内。要求各实验室披露其氟化物测定方法。将各实验室报告的结果与SABS实验室报告的结果进行比较。氟化物浓度为0.13、0.22和0.58毫克/升时,分别有2个、6个和3个实验室能将结果重现至误差在0.05毫克/升以内。在1.1和1.5毫克/升浓度时,没有实验室能达到这种准确度。使用了4种不同的氟化物测定方法。目前,测定氟化物浓度的实验室不够准确,无法确保水氟化过程的安全性。在水氟化开始之前,实验室必须检查其程序以确保获得更好的结果。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验