Payne Marielle, Georgescu Dan, Waite Aaron N, Olson Randall J
Department of Ophthalmology and Visual Sciences, University of Utah Health Sciences Center, Salt Lake City, Utah 84132, USA.
J Cataract Refract Surg. 2006 Aug;32(8):1374-7. doi: 10.1016/j.jcrs.2006.02.066.
To determine the vacuum pressure generated by 4 phacoemulsification devices measured at the phacoemulsification tip.
University ophthalmology department.
The effective vacuum pressures generated by the Sovereign (AMO), Millennium (Bausch & Lomb), Legacy AdvanTec (Alcon Laboratories), and Infiniti (Alcon Laboratories) phacoemulsification machines were measured with a device that isolated the phacoemulsification tip in a chamber connected to a pressure gauge. The 4 machines were tested at multiple vacuum limit settings, and the values were recorded after the foot pedal was fully depressed and the pressure had stabilized. The AdvanTec and Infiniti machines were tested with and without occlusion of the Aspiration Bypass System (ABS) side port (Alcon Laboratories). The Millennium machine was tested using venturi and peristaltic pumps.
The machines generated pressures close to the expected at maximum vacuum settings between 100 mm Hg and 500 mm Hg with few intermachine variations. There was no significant difference between pressures generated using 19- or 20-gauge tips (Millennium and Sovereign). The addition of an ABS side port decreased vacuum by a mean of 12.1% (P < .0001).
Although there were some variations in vacuum pressures among phacoemulsification machines, particularly when an aspiration bypass tip was used, these discrepancies are probably not clinically significant.
测定4种超声乳化设备在超声乳化探头处产生的真空压力。
大学眼科科室。
使用一种将超声乳化探头隔离在与压力计相连的腔室内的设备,测量Sovereign(AMO)、Millennium(博士伦)、Legacy AdvanTec(爱尔康实验室)和Infiniti(爱尔康实验室)超声乳化仪产生的有效真空压力。对这4台仪器在多个真空极限设置下进行测试,在脚踏板完全踩下且压力稳定后记录数值。对AdvanTec和Infiniti仪器在有和无抽吸旁路系统(ABS)侧端口(爱尔康实验室)堵塞的情况下进行测试。使用文丘里泵和蠕动泵对Millennium仪器进行测试。
在100毫米汞柱至500毫米汞柱的最大真空设置下,这些仪器产生的压力接近预期,仪器间差异较小。使用19号或20号探头(Millennium和Sovereign)产生的压力之间无显著差异。添加ABS侧端口使真空度平均降低12.1%(P < .0001)。
尽管超声乳化仪之间的真空压力存在一些差异,特别是在使用抽吸旁路探头时,但这些差异可能在临床上并不显著。