Suppr超能文献

产热:纵向与扭转式超声乳化术

Heat production: Longitudinal versus torsional phacoemulsification.

作者信息

Han Young Keun, Miller Kevin M

机构信息

Department of Ophthalmology, David Geffen School of Medicine at UCLA, and the Jules Stein Eye Institute, Los Angeles, California 90095-7002, USA.

出版信息

J Cataract Refract Surg. 2009 Oct;35(10):1799-805. doi: 10.1016/j.jcrs.2009.04.046.

Abstract

PURPOSE

To compare the heat production of longitudinal versus torsional phacoemulsification under strict laboratory test conditions.

SETTING

Department of Ophthalmology, David Geffen School of Medicine at UCLA, and Jules Stein Eye Institute, Los Angeles, California, USA.

METHODS

Two Infiniti phacoemulsification handpieces were inserted into silicone test chambers filled with a balanced salt solution and imaged serially using a thermal camera. Incision compression was simulated by suspending 25.3 g weights from the silicone chambers. To simulate occlusion of the phacoemulsification tip, the aspiration line was clamped. Peak temperatures were measured 0, 10, 30, 60, and 120 seconds after the commencement of continuous ultrasound power. The 2 handpieces, operating exclusively in longitudinal or torsional modes, were compared 3 ways: (1) using the same power displayed on the instrument console, (2) using identical stroke lengths, and (3) using the same applied energy, a product of stroke length and frequency.

RESULTS

For all 3 comparisons, torsional phacoemulsification resulted in lower temperatures at each time point. At the same displayed power setting, the scenario most familiar to cataract surgeons, longitudinal phacoemulsification elevated temperatures up to 41.5 degrees C more than torsional phacoemulsification.

CONCLUSIONS

Torsional phacoemulsification generated less heat than longitudinal phacoemulsification in all 3 comparison tests. Lower operating temperatures indicate lower heat generation within the same volume of fluid, and this may provide additional thermal protection during cataract surgery.

摘要

目的

在严格的实验室测试条件下比较纵向超声乳化与扭转超声乳化的产热情况。

设置

美国加利福尼亚州洛杉矶市加州大学洛杉矶分校大卫·格芬医学院眼科以及朱尔斯·斯坦眼科研究所。

方法

将两个Infiniti超声乳化手柄插入充满平衡盐溶液的硅胶测试腔中,并用热成像仪进行连续成像。通过在硅胶腔上悬挂25.3克重物来模拟切口压迫。为模拟超声乳化头的堵塞,夹闭吸引管。在连续超声能量开始后0、10、30、6秒和120秒测量峰值温度。对仅以纵向或扭转模式操作的两个手柄进行了三种方式的比较:(1)使用仪器控制台上显示的相同功率;(2)使用相同的冲程长度;(3)使用相同的施加能量,即冲程长度与频率的乘积。

结果

在所有三种比较中,扭转超声乳化在每个时间点产生的温度都较低。在相同的显示功率设置下,这是白内障手术医生最熟悉 的情况,纵向超声乳化产生的温度比扭转超声乳化高出多达41.5摄氏度。

结论

在所有三项比较测试中,扭转超声乳化产生的热量均低于纵向超声乳化。较低的操作温度表明在相同体积的液体中产生的热量较少,这可能在白内障手术期间提供额外的热保护。

相似文献

1
Heat production: Longitudinal versus torsional phacoemulsification.产热:纵向与扭转式超声乳化术
J Cataract Refract Surg. 2009 Oct;35(10):1799-805. doi: 10.1016/j.jcrs.2009.04.046.
8
Thermal study of bare tips with various system parameters and incision sizes.
J Cataract Refract Surg. 2006 May;32(5):867-72. doi: 10.1016/j.jcrs.2005.06.054.

引用本文的文献

9
Heat Generation and Efficiency of a New Modified Phaco Tip and Sleeve.新型改良超声乳化针头及套管的产热与效率
PLoS One. 2016 Aug 3;11(8):e0159049. doi: 10.1371/journal.pone.0159049. eCollection 2016.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验