Gambichler Thilo, Boms Stefanie, Stücker Markus, Kreuter Alexander, Sand Michael, Moussa Georg, Altmeyer Peter, Hoffmann Klaus
Ruhr-University Bochum, Department of Dermatology, Gudrunstr. 56, 44791 Bochum, Germany.
J Biomed Opt. 2005 Jul-Aug;10(4):44008. doi: 10.1117/1.2039086.
There is a lack of systematic investigations comparing optical coherence tomography (OCT) with histology. OCT assessments were performed on the upper back of 16 healthy subjects. Epidermis thickness (ET) was assessed using three methods: first, peak-to-valley analysis of the A-scan (ET-OCT-V); second, manual measurements in the OCT images (ET-OCT-M); third, light microscopic determination using routine histology (ET-Histo). The relationship between the different methods was assessed by means of the Pearson correlation procedure and Bland and Altman plots. We observed a strong correlation between ET-Histo (79.4+/-21.9 microm) and ET-OCT-V (79.2+/-15.5 microm, r=0.77) and ET-OCT-M (82.9+/-15.8 microm, r=0.75), respectively. Bland and Altman plots revealed a bias of -0.19 microm (95% limits of agreement: -27.94 microm to 27.56 microm) for ET-OCT-V versus ET-Histo and a bias of 3.44 microm (95% limits of agreement: -24.9 microm to 31.78 microm) for ET-OCT-M versus ET-Histo. Despite the strong correlation and low bias observed, the 95% limits of agreement demonstrated an unsatisfactory numerical agreement between the two OCT methods and routine histology indicating that these methods cannot be employed interchangeably. Regarding practicability, precision, and indication spectrum, ET-OCT-V and ET-OCT-M are of different clinical value.
缺乏将光学相干断层扫描(OCT)与组织学进行比较的系统性研究。对16名健康受试者的上背部进行了OCT评估。使用三种方法评估表皮厚度(ET):第一,A扫描的峰谷分析(ET-OCT-V);第二,在OCT图像中进行手动测量(ET-OCT-M);第三,使用常规组织学进行光学显微镜测定(ET-Histo)。通过Pearson相关程序以及Bland和Altman图评估不同方法之间的关系。我们分别观察到ET-Histo(79.4±21.9微米)与ET-OCT-V(79.2±15.5微米,r = 0.77)以及ET-Histo与ET-OCT-M(82.9±15.8微米,r = 0.75)之间存在强相关性。Bland和Altman图显示,ET-OCT-V与ET-Histo相比偏差为-0.19微米(95%一致性界限:-27.94微米至27.56微米),ET-OCT-M与ET-Histo相比偏差为3.44微米(95%一致性界限:-24.9微米至31.78微米)。尽管观察到强相关性和低偏差,但95%一致性界限表明两种OCT方法与常规组织学之间的数值一致性并不理想,这表明这些方法不能互换使用。在实用性、精确性和适应症范围方面,ET-OCT-V和ET-OCT-M具有不同的临床价值。