Hochrainer Dieter, Hölz Hubert, Kreher Christoph, Scaffidi Luigi, Spallek Michael, Wachtel Herbert
Boehringer Ingelheim Pharma GmbH and Co. KG, Ingelheim, Germany.
J Aerosol Med. 2005 Fall;18(3):273-82. doi: 10.1089/jam.2005.18.273.
Apart from particle size distribution, spray velocity is one of the most important aerosol characteristics that influence lung deposition of inhaled drugs. The time period over which the aerosol is released (spray duration) is also important for coordination of inhalation. Respimat Soft Mist Inhaler (SMI) is a new generation, propellant-free inhaler that delivers drug to the lung much more efficiently than pressurised metered dose inhalers (pMDIs). The objective of this study was to compare the velocity and spray duration of aerosol clouds produced by Respimat SMI with those from a variety of chlorofluorocarbon (CFC) and hydrofluoroalkane (HFA) pMDIs. All inhalers contained solutions or suspensions of bronchodilators. A videorecording method was used to determine the aerosol velocity. For spray duration, the time for generation of the Soft Mist by Respimat SMI was initially determined using three different methods (videorecording [techniques A and B], laser light diffraction and rotating disc). Videorecording was then used to compare the spray duration of Respimat SMI with those from the other inhalers. The Soft Mist produced by Respimat SMI moved much more slowly and had a more prolonged duration than aerosol clouds from pMDIs (mean velocity at a 10-cm distance from the nozzle: Respimat SMI, 0.8 m/sec; pMDIs, 2.0-8.4 m/sec; mean duration: Respimat SMI, 1.5 sec; pMDIs, 0.15-0.36 sec). These characteristics should result in improved lung and reduced oropharyngeal deposition, and are likely to simplify coordination of inhaler actuation and inhalation compared with pMDIs.
除了粒径分布外,喷雾速度是影响吸入药物肺部沉积的最重要的气溶胶特性之一。气溶胶释放的时间段(喷雾持续时间)对于吸入的协调性也很重要。Respimat软雾吸入器(SMI)是新一代不含推进剂的吸入器,其将药物输送到肺部的效率比压力定量吸入器(pMDIs)高得多。本研究的目的是比较Respimat SMI产生的气溶胶云与各种氯氟烃(CFC)和氢氟烷烃(HFA)pMDIs产生的气溶胶云的速度和喷雾持续时间。所有吸入器均含有支气管扩张剂的溶液或混悬液。采用录像方法测定气溶胶速度。对于喷雾持续时间,最初使用三种不同方法(录像[技术A和B]、激光衍射和转盘)确定Respimat SMI产生软雾的时间。然后使用录像比较Respimat SMI与其他吸入器的喷雾持续时间。Respimat SMI产生的软雾移动速度慢得多,持续时间比pMDIs产生的气溶胶云更长(距喷嘴10 cm处的平均速度:Respimat SMI,0.8 m/秒;pMDIs,2.0 - 8.4 m/秒;平均持续时间:Respimat SMI,1.5秒;pMDIs,0.15 - 0.36秒)。这些特性应能改善肺部沉积并减少口咽部沉积,与pMDIs相比,可能会简化吸入器启动和吸入的协调性。