Ng R H, Sparks K M, Statland B E
Department of Pathology & Laboratory Medicine, Methodist Hospital of Indiana, Indianapolis 46202.
Clin Chem. 1992 Jul;38(7):1371-2.
We evaluated a colorimetric assay of potassium in plasma and serum with the Boehringer Mannheim Reflotron reflectance photometric analyzer, which is designed for near-patient testing in hospitals and physicians' offices. This potassium method does not require calibration or instrument maintenance by the operator. Analysis of 30 microL of plasma or serum takes approximately 140 s. Within-day imprecision (CV) was 1.0-1.2%. Total CVs over a 1-month period were 1.0-1.4%. Patients' results from the Reflotron correlated well with those from the IL 643 flame photometer and the Beckman Synchron CX3 ion-selective electrode methods. The accuracy of Reflotron values was also verified with Standard Reference Material 956 from the National Institute of Standards and Technology.
我们使用博林格曼海姆Reflotron反射光度分析仪评估了血浆和血清中钾的比色测定法,该分析仪专为医院和医生办公室的即时检测而设计。这种钾检测方法不需要操作人员进行校准或仪器维护。分析30微升血浆或血清大约需要140秒。日内不精密度(CV)为1.0 - 1.2%。1个月期间的总CV为1.0 - 1.4%。Reflotron检测的患者结果与IL 643火焰光度计和贝克曼Synchron CX3离子选择电极法的结果相关性良好。Reflotron值的准确性也通过美国国家标准与技术研究院的标准参考物质956得到了验证。