Ng R H, Sparks K M, Statland B E
Department of Pathology & Laboratory Medicine, Methodist Hospital of Indiana, Indianapolis 46202.
Clin Chem. 1991 Mar;37(3):435-7.
We evaluated the quantification of high-density lipoprotein (HDL) cholesterol in plasma with the Boehringer Mannheim Reflotron reflectance photometric analyzer. The Reflotron is designed for testing in small to medium-size laboratories and physicians' offices. This HDL method does not require a manual precipitation step because the reagent tab contains dextran sulfate (Mr 50,000) and magnesium acetate. It takes 90 s to complete an analysis of 30 microL of plasma. Within-day standard deviations (SDs) were 0.02-0.04 mmol/L (6-16 mg/L). Total SDs over a three-month period were 0.03-0.06 mmol/L (11-23 mg/L). The Reflotron values averaged 0.02 mmol/L (6 mg/L) or 1.3% lower than the Hitachi 737 values; the standard error of the estimate (Sy.x) was 0.07 mmol/L (29 mg/L).
我们使用勃林格殷格翰Reflotron反射光度分析仪评估了血浆中高密度脂蛋白(HDL)胆固醇的定量分析。Reflotron专为中小型实验室和医生办公室的检测而设计。这种HDL检测方法不需要手动沉淀步骤,因为试剂条中含有硫酸葡聚糖(分子量50,000)和乙酸镁。分析30微升血浆需要90秒。日内标准差(SDs)为0.02 - 0.04毫摩尔/升(6 - 16毫克/升)。三个月期间的总标准差为0.03 - 0.06毫摩尔/升(11 - 23毫克/升)。Reflotron测得的值平均比日立737的值低0.02毫摩尔/升(6毫克/升)或1.3%;估计标准误差(Sy.x)为0.07毫摩尔/升(29毫克/升)。