使用新型试剂载体(Reflotron)进行血浆钾测量:与离子选择性电极结果的比较。

Plasma potassium measurement with a new reagent carrier (Reflotron): comparison with ion-selective electrode results.

作者信息

Braun J P, Carstensen C A

机构信息

Département de Physiopathologie, Ecole Nationale Vétérinaire, Toulouse, France.

出版信息

Res Vet Sci. 1993 Jan;54(1):127-9. doi: 10.1016/0034-5288(93)90022-8.

Abstract

Potassium concentrations were measured in the plasma of 336 animals with a new reagent carrier (Reflotron; Boehringer Mannheim) K+ and with an ion-specific electrode system: results were highly correlated (r = 0.991; y = 0.993 x + 0.02) and day-to-day coefficient of variation of the new reagent measurements was lower than 2.5 per cent. This system offers a good alternative to the ion-selective electrode system for plasma potassium measurement in veterinary practice.

摘要

使用一种新型试剂载体(Reflotron;勃林格殷格翰公司)K⁺和离子特异性电极系统对336只动物的血浆中的钾浓度进行了测量:结果高度相关(r = 0.991;y = 0.993x + 0.02),并且新型试剂测量的每日变异系数低于2.5%。该系统为兽医实践中血浆钾测量的离子选择性电极系统提供了一个很好的替代方案。