使用微尖端和外部水压传感器获得的测量值的多中心比较。

A multicentered comparison of measurements obtained with microtip and external water pressure transducers.

作者信息

Hundley Andrew F, Brown Morton B, Brubaker Linda, Cundiff Geoffrey W, Kreder Karl, Lotze Peter, Richter Holly E, Zyczynski Halina, Weber Anne M, Visco Anthony G

机构信息

Division of Female Pelvic Medicine and Reconstructive Pelvic Surgery, University of North Carolina at Chapel Hill, Chapel Hill, NC, USA.

出版信息

Int Urogynecol J Pelvic Floor Dysfunct. 2006 Jun;17(4):400-6. doi: 10.1007/s00192-005-0027-0. Epub 2005 Nov 12.

Abstract

This study compared simultaneous intravesical pressure readings obtained with catheter-mounted microtip transducers and external water pressure transducer catheters during filling cystometry. Women undergoing multichannel urodynamic testing were randomly assigned to one of three groups: two microtip catheters, two external water pressure transducer catheters, or one of each type. Intravesical pressure was measured simultaneously for each transducer combination in each subject for minimal and maximal Valsalva effort and minimal, moderate, and maximal cough effort at two sequential bladder volumes (150 and 300 ml). Paired t tests were used to compare the means of the intravesical pressure obtained by the two types of catheters. The largest mean differences were observed when comparing microtip and water pressure transducers. Correlations of maximum pressure were consistently high between two microtip transducers and two water pressure transducers but lower for the microtip-water combination. Excellent reproducibility was demonstrated with transducers of similar types for intravesical pressures recorded during Valsalva and cough in women without prolapse. However, considerable variability was seen in pressures recorded by different transducers, particularly dependent on the water catheter manufacturer, indicating that intravesical pressure recordings from microtip and water-based systems are not interchangeable.

摘要

本研究比较了在膀胱测压充盈期使用导管安装式微尖端换能器和外置水压力换能器导管同时获得的膀胱内压读数。接受多通道尿动力学检测的女性被随机分为三组之一:两根微尖端导管组、两根外置水压力换能器导管组或每种类型各一根导管组。在每个受试者中,针对每种换能器组合,在两个连续的膀胱容量(150毫升和300毫升)下,同时测量最小和最大瓦尔萨尔瓦动作以及最小、中度和最大咳嗽动作时的膀胱内压。采用配对t检验比较两种导管获得的膀胱内压平均值。在比较微尖端换能器和水压力换能器时观察到最大的平均差异。两根微尖端换能器之间以及两根水压力换能器之间的最大压力相关性始终很高,但微尖端 - 水压力组合的相关性较低。对于无子宫脱垂女性在瓦尔萨尔瓦动作和咳嗽期间记录的膀胱内压,同类换能器显示出极佳的可重复性。然而,不同换能器记录的压力存在相当大的变异性,特别是取决于水导管制造商,这表明微尖端系统和水基系统记录的膀胱内压不可互换。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验