The present study evaluated the applicability of a rebound tonometer in measuring intraocular pressure (IOP) in rats and mice.
The accuracy of the TonoLab rebound tonometer was determined in cannulated mouse and rat eyes. IOP was manipulated by changing reservoir height, and tonometer pressure readings were recorded by an independent observer. IOP values were recorded in conscious Wistar rats and in four different strains of mice. The effects of anesthesia on IOP were evaluated in two different strains of mice.
The IOP readings generated by the rebound tonometer correlated very well with the actual pressure in the eye. In rats, this linear correlation had a slope of 0.96 +/- 0.05 (mean +/- SEM, n = 4) and a Y-intercept of -2.1 +/- 1.2. In mice, the slope was 0.99 +/- 0.05 (n = 3), and the Y-intercept was 0.8 +/- 1.4. Using this method, the resting IOP of conscious male Wistar rats was observed to be 18.4 +/- 0.1 mm Hg (n = 132). In mice, strain differences in IOP were detected. Baseline IOP values in Balb/c, C57-BL/6, CBA, and 11- to 12-month-old DBA/2J mice were 10.6 +/- 0.6, 13.3 +/- 0.3, 16.4 +/- 0.3, and 19.3 +/- 0.4 mm Hg (n = 12), respectively. In separated studies, anesthesia lowered IOP from 14.3 +/- 0.9 to 9.2 +/- 0.5 mm Hg (n = 8) in C57-BL/6 mice, and from 16.6 +/- 0.4 to 9.4 +/- 0.6 mm Hg (n = 10) in CBA mice.
The rebound tonometer was easy to use and accurately measured IOP in rats and mice. This technique, together with advances in genetic and other biological studies in rodents, will be valuable in the further understanding of the etiology and pathology of glaucoma.
本研究评估了回弹式眼压计在测量大鼠和小鼠眼压(IOP)方面的适用性。
在插管的小鼠和大鼠眼睛中测定TonoLab回弹式眼压计的准确性。通过改变储液器高度来控制眼压,并由一名独立观察者记录眼压计的压力读数。在清醒的Wistar大鼠和四种不同品系的小鼠中记录眼压值。在两种不同品系的小鼠中评估麻醉对眼压的影响。
回弹式眼压计测得的眼压读数与眼内实际压力相关性非常好。在大鼠中,这种线性相关性的斜率为0.96±0.05(平均值±标准误,n = 4),Y轴截距为-2.1±1.2。在小鼠中,斜率为0.99±0.05(n = 3),Y轴截距为0.8±1.4。使用这种方法,观察到清醒雄性Wistar大鼠的静息眼压为18.4±0.1 mmHg(n = 132)。在小鼠中,检测到眼压的品系差异。Balb/c、C57-BL/6、CBA和11至12月龄DBA/2J小鼠的基线眼压值分别为10.6±0.6、13.3±0.3、16.4±0.3和19.3±0.4 mmHg(n = 12)。在单独的研究中,麻醉使C57-BL/6小鼠的眼压从14.3±0.9 mmHg降至9.2±0.5 mmHg(n = 8),使CBA小鼠的眼压从16.6±0.4 mmHg降至9.4±0.6 mmHg(n = 10)。
回弹式眼压计易于使用,能准确测量大鼠和小鼠的眼压。这项技术,连同啮齿动物遗传和其他生物学研究的进展,将有助于进一步了解青光眼的病因和病理。