Suppr超能文献

用于对比剂增强磁共振成像的脂肪抑制三维双回波狄克逊技术。

Fat-suppressed three-dimensional dual echo Dixon technique for contrast agent enhanced MRI.

作者信息

Ma Jingfei, Vu Anthony T, Son Jong Bum, Choi Haesun, Hazle John D

机构信息

Department of Imaging Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas 77030, USA.

出版信息

J Magn Reson Imaging. 2006 Jan;23(1):36-41. doi: 10.1002/jmri.20470.

Abstract

PURPOSE

To develop a fast T1-weighted, fat-suppressed three-dimensional dual echo Dixon technique and to demonstrate its use in contrast agent enhanced MRI.

MATERIALS AND METHODS

A product fast three-dimensional gradient echo pulse sequence was modified to acquire dual echoes after each RF excitation with water and fat signals in-phase (IP) and opposed-phase (OP), respectively. An on-line reconstruction algorithm was implemented to automatically generate separate water and fat images. The signal to noise ratio (SNR) of the new technique was compared to that of the product technique in phantom. In vivo abdomen and breast images of cancer patients were acquired at 1.5 Tesla using both techniques before and after intravenous administration of gadolinium contrast agent.

RESULTS

In phantom, the new technique yields a close to the theoretically predicted 41% increase in SNR in comparison to the product technique without fat suppression (FS). In vivo images of the new technique show noticeably improved FS and image quality in comparison to the images acquired of the same patients using the product technique with FS.

CONCLUSION

The three-dimensional dual echo Dixon technique provides excellent image quality and can be used for T1-weighted, fat-suppressed imaging with contrast agent injection.

摘要

目的

开发一种快速T1加权、脂肪抑制的三维双回波狄克逊技术,并展示其在对比剂增强磁共振成像中的应用。

材料与方法

对一种商用快速三维梯度回波脉冲序列进行修改,以便在每次射频激发后分别采集水和脂肪信号同相(IP)和反相(OP)的双回波。实施一种在线重建算法以自动生成单独的水和脂肪图像。在体模中将新技术的信噪比(SNR)与商用技术的信噪比进行比较。在静脉注射钆对比剂之前和之后,使用这两种技术在1.5特斯拉下采集癌症患者的体内腹部和乳腺图像。

结果

在体模中,与无脂肪抑制(FS)的商用技术相比,新技术的SNR提高了接近理论预测的41%。与使用具有FS的商用技术为同一患者采集的图像相比,新技术的体内图像显示出明显改善的FS和图像质量。

结论

三维双回波狄克逊技术提供了优异的图像质量,可用于注射对比剂后的T1加权、脂肪抑制成像。